Get 20M+ Full-Text Papers For Less Than $1.50/day. Start a 7-Day Trial for You or Your Team.

Learn More →

ISCEV Standard for full-field clinical electroretinography (2008 update)

ISCEV Standard for full-field clinical electroretinography (2008 update) This document, from the International Society for Clinical Electrophysiology of Vision (ISCEV), presents an updated and revised ISCEV Standard for clinical electroretinography (ERG). The parameters for flash stimulation and background adaptation have been tightened, and responses renamed to indicate the flash strength (in cd·s·m−2). The ISCEV Standard specifies five responses: (1) Dark-adapted 0.01 ERG (rod response); (2) Dark-adapted 3.0 ERG (combined rod–cone response); (3) Dark-adapted 3.0 oscillatory potentials; (4) Light-adapted 3.0 ERG (cone response); (5) Light-adapted 3.0 flicker (30 Hz flicker). An additional Dark-adapted 10.0 ERG or Dark-adapted 30.0 ERG response is recommended. http://www.deepdyve.com/assets/images/DeepDyve-Logo-lg.png Documenta Ophthalmologica Springer Journals

ISCEV Standard for full-field clinical electroretinography (2008 update)

Loading next page...
 
/lp/springer-journals/iscev-standard-for-full-field-clinical-electroretinography-2008-update-g5jDBMzfsf

References (12)

Publisher
Springer Journals
Copyright
Copyright © 2008 by Springer-Verlag
Subject
Medicine & Public Health; Ophthalmology
ISSN
0012-4486
eISSN
1573-2622
DOI
10.1007/s10633-008-9155-4
pmid
19030905
Publisher site
See Article on Publisher Site

Abstract

This document, from the International Society for Clinical Electrophysiology of Vision (ISCEV), presents an updated and revised ISCEV Standard for clinical electroretinography (ERG). The parameters for flash stimulation and background adaptation have been tightened, and responses renamed to indicate the flash strength (in cd·s·m−2). The ISCEV Standard specifies five responses: (1) Dark-adapted 0.01 ERG (rod response); (2) Dark-adapted 3.0 ERG (combined rod–cone response); (3) Dark-adapted 3.0 oscillatory potentials; (4) Light-adapted 3.0 ERG (cone response); (5) Light-adapted 3.0 flicker (30 Hz flicker). An additional Dark-adapted 10.0 ERG or Dark-adapted 30.0 ERG response is recommended.

Journal

Documenta OphthalmologicaSpringer Journals

Published: Nov 22, 2008

There are no references for this article.