Effects of proximity and noise level of phased array coil elements on overall signal-to-noise in parallel MR spectroscopy

Effects of proximity and noise level of phased array coil elements on overall signal-to-noise in... Parallel imaging using phased array coils facilitates accelerated magnetic resonance imaging (MRI) and spectroscopy (MRS). Parallel data reconstruction requires the combination of data from individual coil elements, but limited combination algorithms currently exist for higher-order phased arrays and MRS data. Here, we present a systematic framework for identifying coil proximity-related signal inhomogeneities and noise levels in phased array coils that may affect sensitivity of parallel MRS. Single-voxel MRS was acquired in nine voxel positions in a brain spectroscopy phantom on a 3T whole-body MR scanner using commercially available 64-, 32-, and 20-channel phased array coils. Spectra produced by individual coil elements were combined using both a signal-to-noise ratio (SNR) threshold and based on the position of individual coil elements. SNR and metabolite Cramer-Rao lower bounds (CRLBs) from the final combined spectra were used as metrics to compare combination strategies and the effects of the phased array geometry and individual coil proximity. Comparisons were performed using one-way repeated measures ANOVA and post-hoc Tukey's range test (p<0.05). The 32-channel phased array coil produced the highest overall SNR compared to the 64-channel (p=0.0009) or 20-channel coils (p=0.003). Low SNR spectra from individual coil elements in the 64-channel coil can reduce the overall SNR when simply combining spectra from all elements. SNR varied significantly as a function of voxel position (F=58.3, p<0.0001) and SNR threshold for all phased arrays (p<0.05 for 64-, 32-, and 20-channel coils). Metabolite CRLBs were dependent on the combination strategy. We demonstrate the importance of the sampling voxel position and coil proximity on overall SNR in parallel MRS data acquisition, with significant SNR improvements after selectively filtering individual spectra based on pre-determined SNR thresholds which must be optimized for each phased array coil element and volume of interest. http://www.deepdyve.com/assets/images/DeepDyve-Logo-lg.png Magnetic Resonance Imaging Elsevier

Effects of proximity and noise level of phased array coil elements on overall signal-to-noise in parallel MR spectroscopy

Loading next page...
 
/lp/elsevier/effects-of-proximity-and-noise-level-of-phased-array-coil-elements-on-KYaU23LIX0
Publisher
Elsevier
Copyright
Copyright © 2017 Elsevier Inc.
ISSN
0730-725X
D.O.I.
10.1016/j.mri.2017.12.001
Publisher site
See Article on Publisher Site

Abstract

Parallel imaging using phased array coils facilitates accelerated magnetic resonance imaging (MRI) and spectroscopy (MRS). Parallel data reconstruction requires the combination of data from individual coil elements, but limited combination algorithms currently exist for higher-order phased arrays and MRS data. Here, we present a systematic framework for identifying coil proximity-related signal inhomogeneities and noise levels in phased array coils that may affect sensitivity of parallel MRS. Single-voxel MRS was acquired in nine voxel positions in a brain spectroscopy phantom on a 3T whole-body MR scanner using commercially available 64-, 32-, and 20-channel phased array coils. Spectra produced by individual coil elements were combined using both a signal-to-noise ratio (SNR) threshold and based on the position of individual coil elements. SNR and metabolite Cramer-Rao lower bounds (CRLBs) from the final combined spectra were used as metrics to compare combination strategies and the effects of the phased array geometry and individual coil proximity. Comparisons were performed using one-way repeated measures ANOVA and post-hoc Tukey's range test (p<0.05). The 32-channel phased array coil produced the highest overall SNR compared to the 64-channel (p=0.0009) or 20-channel coils (p=0.003). Low SNR spectra from individual coil elements in the 64-channel coil can reduce the overall SNR when simply combining spectra from all elements. SNR varied significantly as a function of voxel position (F=58.3, p<0.0001) and SNR threshold for all phased arrays (p<0.05 for 64-, 32-, and 20-channel coils). Metabolite CRLBs were dependent on the combination strategy. We demonstrate the importance of the sampling voxel position and coil proximity on overall SNR in parallel MRS data acquisition, with significant SNR improvements after selectively filtering individual spectra based on pre-determined SNR thresholds which must be optimized for each phased array coil element and volume of interest.

Journal

Magnetic Resonance ImagingElsevier

Published: Apr 1, 2018

References

You’re reading a free preview. Subscribe to read the entire article.


DeepDyve is your
personal research library

It’s your single place to instantly
discover and read the research
that matters to you.

Enjoy affordable access to
over 18 million articles from more than
15,000 peer-reviewed journals.

All for just $49/month

Explore the DeepDyve Library

Search

Query the DeepDyve database, plus search all of PubMed and Google Scholar seamlessly

Organize

Save any article or search result from DeepDyve, PubMed, and Google Scholar... all in one place.

Access

Get unlimited, online access to over 18 million full-text articles from more than 15,000 scientific journals.

Your journals are on DeepDyve

Read from thousands of the leading scholarly journals from SpringerNature, Elsevier, Wiley-Blackwell, Oxford University Press and more.

All the latest content is available, no embargo periods.

See the journals in your area

DeepDyve

Freelancer

DeepDyve

Pro

Price

FREE

$49/month
$360/year

Save searches from
Google Scholar,
PubMed

Create lists to
organize your research

Export lists, citations

Read DeepDyve articles

Abstract access only

Unlimited access to over
18 million full-text articles

Print

20 pages / month

PDF Discount

20% off