Testing and validating a homogeneous immunometric assay for interference Abstract We analyzed 95 sera, demonstrating interference in a previous study, with the Kryptor homogeneous time-resolved fluorescence resonance energy transfer carcinoembryonic antigen (CEA) immunoassay (Brahms AG, Berlin, Germany). Only one serum differed, i.e., 6.0 μg/l for Kryptor vs. 13.3 μg/l for a microtiter plate in-house immunofluorometric assay (IFMA), using both aggregated mouse immunoglobulins as blocker and capture monoclonal antibody (Mab) F(ab′) 2 -fragments to reduce interference. Sera were further analyzed with experimental IFMAs with intact capture Mabs and without blocker. The Kryptor-CEA assay Mab GFR44 (capture) had elevated results in 71% of sera with the Kryptor Mab G15 (tracer) or in 81% with our tracer Mab. G15 (capture) had 65% elevated results with GFR44 (tracer) or 99% with our tracer Mab. The latter was reduced from 99% to 62% by adding Kryptor blocker to the buffer or to 30% by adding our blocker. Finally, combining both assay Mabs and assay buffer from Kryptor still gave interference in 16% of sera. Kryptor-CEA assay results thus agreed with our inhouse CEA assay results, showing no interference. As the Kryptor-CEA assay antibodies were sensitive to interference and the Kryptor-CEA assay buffer did not reduce interference as efficiently as our in-house assay buffer, the Kryptor-CEA assay format was crucial for the absence of interference. http://www.deepdyve.com/assets/images/DeepDyve-Logo-lg.png Clinical Chemistry and Laboratory Medicine (CCLM) de Gruyter

Testing and validating a homogeneous immunometric assay for interference

Loading next page...
 
/lp/de-gruyter/testing-and-validating-a-homogeneous-immunometric-assay-for-QXwn4CNyPi

References (26)

Publisher
de Gruyter
Copyright
Copyright © 2004 by the
ISSN
1434-6621
eISSN
1437-4331
DOI
10.1515/CCLM.2004.038
pmid
15061363
Publisher site
See Article on Publisher Site

Abstract

Abstract We analyzed 95 sera, demonstrating interference in a previous study, with the Kryptor homogeneous time-resolved fluorescence resonance energy transfer carcinoembryonic antigen (CEA) immunoassay (Brahms AG, Berlin, Germany). Only one serum differed, i.e., 6.0 μg/l for Kryptor vs. 13.3 μg/l for a microtiter plate in-house immunofluorometric assay (IFMA), using both aggregated mouse immunoglobulins as blocker and capture monoclonal antibody (Mab) F(ab′) 2 -fragments to reduce interference. Sera were further analyzed with experimental IFMAs with intact capture Mabs and without blocker. The Kryptor-CEA assay Mab GFR44 (capture) had elevated results in 71% of sera with the Kryptor Mab G15 (tracer) or in 81% with our tracer Mab. G15 (capture) had 65% elevated results with GFR44 (tracer) or 99% with our tracer Mab. The latter was reduced from 99% to 62% by adding Kryptor blocker to the buffer or to 30% by adding our blocker. Finally, combining both assay Mabs and assay buffer from Kryptor still gave interference in 16% of sera. Kryptor-CEA assay results thus agreed with our inhouse CEA assay results, showing no interference. As the Kryptor-CEA assay antibodies were sensitive to interference and the Kryptor-CEA assay buffer did not reduce interference as efficiently as our in-house assay buffer, the Kryptor-CEA assay format was crucial for the absence of interference.

Journal

Clinical Chemistry and Laboratory Medicine (CCLM)de Gruyter

Published: Feb 18, 2004

There are no references for this article.