Initial experience with imaging of the lower extremity arteries in an open 1.0 Tesla MRI system using the triggered angiography non-contrast-enhanced sequence (TRANCE) compared to digital subtraction angiography (DSA)

Initial experience with imaging of the lower extremity arteries in an open 1.0 Tesla MRI system... AbstractPurpose:The aim of this study was to evaluate the feasibility and validity of arterial lower limb imaging with triggered angiography non-contrast enhanced (TRANCE) in an open MRI at 1.0 Tesla (T) compared to digital subtraction angiography (DSA).Material and methods:ECG-gated, non-contrast-enhanced magnetic resonance angiography (MRA) was performed in a 1.0-T high-field open magnetic resonance imaging (MRI) system which generates a vertical magnetic field. Three acquisition levels were defined (abdominal and pelvic level, arterial segments above the knee and segments below the knee) and a total of 1782 vessel diameter measurements were taken on a total of 11 patients with suspected peripheral arterial occlusive disease (PAOD) (8 men, 3 women; average age 66 years). In each patient, 162 vessel segments (81 each with TRANCE and DSA) were defined and measured. Pearson correlation coefficients were calculated.Results:At the abdominal/pelvic level, all mean values measured with DSA exceeded the mean values obtained with TRANCE. Above the knee, mean vessel diameters were measured smaller in DSA in six, equal in three, and larger in two vessel segments. Below the knee, all measured averages, except for the tibiofibular tract (TFT) measurements, were larger in TRANCE. In total, two small (≤0.3), two moderate (>0.3), 11 good (>0.5), 10 high (>0.7) and 13 very high (>0.8) correlations were obtained.Conclusions:Non-contrast-enhanced imaging of the lower limb arteries using a TRANCE-sequence in a 1.0 T open MRI system is feasible with the protocol presented; however, TRANCE tends to underestimate larger vessels and overestimate smaller vessels compared to DSA. http://www.deepdyve.com/assets/images/DeepDyve-Logo-lg.png Biomedical Engineering / Biomedizinische Technik de Gruyter

Initial experience with imaging of the lower extremity arteries in an open 1.0 Tesla MRI system using the triggered angiography non-contrast-enhanced sequence (TRANCE) compared to digital subtraction angiography (DSA)

Loading next page...
 
/lp/degruyter/initial-experience-with-imaging-of-the-lower-extremity-arteries-in-an-Yp6e1E9JQr
Publisher
De Gruyter
Copyright
©2016 Walter de Gruyter GmbH, Berlin/Boston
ISSN
1862-278X
eISSN
1862-278X
D.O.I.
10.1515/bmt-2014-0181
Publisher site
See Article on Publisher Site

Abstract

AbstractPurpose:The aim of this study was to evaluate the feasibility and validity of arterial lower limb imaging with triggered angiography non-contrast enhanced (TRANCE) in an open MRI at 1.0 Tesla (T) compared to digital subtraction angiography (DSA).Material and methods:ECG-gated, non-contrast-enhanced magnetic resonance angiography (MRA) was performed in a 1.0-T high-field open magnetic resonance imaging (MRI) system which generates a vertical magnetic field. Three acquisition levels were defined (abdominal and pelvic level, arterial segments above the knee and segments below the knee) and a total of 1782 vessel diameter measurements were taken on a total of 11 patients with suspected peripheral arterial occlusive disease (PAOD) (8 men, 3 women; average age 66 years). In each patient, 162 vessel segments (81 each with TRANCE and DSA) were defined and measured. Pearson correlation coefficients were calculated.Results:At the abdominal/pelvic level, all mean values measured with DSA exceeded the mean values obtained with TRANCE. Above the knee, mean vessel diameters were measured smaller in DSA in six, equal in three, and larger in two vessel segments. Below the knee, all measured averages, except for the tibiofibular tract (TFT) measurements, were larger in TRANCE. In total, two small (≤0.3), two moderate (>0.3), 11 good (>0.5), 10 high (>0.7) and 13 very high (>0.8) correlations were obtained.Conclusions:Non-contrast-enhanced imaging of the lower limb arteries using a TRANCE-sequence in a 1.0 T open MRI system is feasible with the protocol presented; however, TRANCE tends to underestimate larger vessels and overestimate smaller vessels compared to DSA.

Journal

Biomedical Engineering / Biomedizinische Technikde Gruyter

Published: Aug 1, 2016

There are no references for this article.

You’re reading a free preview. Subscribe to read the entire article.


DeepDyve is your
personal research library

It’s your single place to instantly
discover and read the research
that matters to you.

Enjoy affordable access to
over 18 million articles from more than
15,000 peer-reviewed journals.

All for just $49/month

Explore the DeepDyve Library

Search

Query the DeepDyve database, plus search all of PubMed and Google Scholar seamlessly

Organize

Save any article or search result from DeepDyve, PubMed, and Google Scholar... all in one place.

Access

Get unlimited, online access to over 18 million full-text articles from more than 15,000 scientific journals.

Your journals are on DeepDyve

Read from thousands of the leading scholarly journals from SpringerNature, Elsevier, Wiley-Blackwell, Oxford University Press and more.

All the latest content is available, no embargo periods.

See the journals in your area

DeepDyve

Freelancer

DeepDyve

Pro

Price

FREE

$49/month
$360/year

Save searches from
Google Scholar,
PubMed

Create lists to
organize your research

Export lists, citations

Read DeepDyve articles

Abstract access only

Unlimited access to over
18 million full-text articles

Print

20 pages / month

PDF Discount

20% off