Real-time scintillation array dosimetry for radiotherapy: The advantages of photomultiplier detectors

Real-time scintillation array dosimetry for radiotherapy: The advantages of photomultiplier... Purpose: In this paper, a photomultiplier tube (PMT) array dosimetry system has been developed and tested for the real-time readout of multiple scintillation signals from fiber optic dosimeters. It provides array dosimetry with the advantages in sensitivity provided by a PMT, but without the need for a separate PMT for each detector element. Methods: The PMT array system consisted of a multianode PMT, a multichannel data acquisition system, housing and optic fiber connections suitable for clinical use. The reproducibility, channel uniformity, channel crosstalk, acquisition speed, and sensitivity of the PMT array were quantified using a constant light source. Its performance was compared to other readout systems used in scintillation dosimetry. An in vivo HDR brachytherapy treatment was used as an example of a clinical application of the dosimetry system to the measurement of dose at multiple sites in the rectum. The PMT array system was also tested in the pulsed beam of a linear accelerator to test its response speed and its application with two separate methods of Cerenkov background removal. Results: The PMT array dosimetry system was highly reproducible with a measurement uncertainty of 0.13% for a 10 s acquisition period. Optical crosstalk between neighboring channels was accounted for by omitting every second channel. A mathematical procedure was used to account for the crosstalk in next-neighbor channels. The speed and sensitivity of the PMT array system were found be superior to CCD cameras, allowing for measurement of more rapid changes in dose rate. This was further demonstrated by measuring the dose delivered by individual photon pulses of a linear accelerator beam. Conclusions: The PMT array system has advantages over CCD camera-based systems for the readout of scintillation light. It provided a more sensitive, more accurate, and faster response to meet the demands of future developments in treatment delivery. http://www.deepdyve.com/assets/images/DeepDyve-Logo-lg.png Medical Physics American Association of Physicists in Medicine

Real-time scintillation array dosimetry for radiotherapy: The advantages of photomultiplier detectors

Loading next page...
 
/lp/american-association-of-physicists-in-medicine/real-time-scintillation-array-dosimetry-for-radiotherapy-the-W7rKxHujH4
Publisher
American Association of Physicists in Medicine
Copyright
Copyright © 2012 American Association of Physicists in Medicine
ISSN
0094-2405
DOI
10.1118/1.3690465
pmid
22482594
Publisher site
See Article on Publisher Site

Abstract

Purpose: In this paper, a photomultiplier tube (PMT) array dosimetry system has been developed and tested for the real-time readout of multiple scintillation signals from fiber optic dosimeters. It provides array dosimetry with the advantages in sensitivity provided by a PMT, but without the need for a separate PMT for each detector element. Methods: The PMT array system consisted of a multianode PMT, a multichannel data acquisition system, housing and optic fiber connections suitable for clinical use. The reproducibility, channel uniformity, channel crosstalk, acquisition speed, and sensitivity of the PMT array were quantified using a constant light source. Its performance was compared to other readout systems used in scintillation dosimetry. An in vivo HDR brachytherapy treatment was used as an example of a clinical application of the dosimetry system to the measurement of dose at multiple sites in the rectum. The PMT array system was also tested in the pulsed beam of a linear accelerator to test its response speed and its application with two separate methods of Cerenkov background removal. Results: The PMT array dosimetry system was highly reproducible with a measurement uncertainty of 0.13% for a 10 s acquisition period. Optical crosstalk between neighboring channels was accounted for by omitting every second channel. A mathematical procedure was used to account for the crosstalk in next-neighbor channels. The speed and sensitivity of the PMT array system were found be superior to CCD cameras, allowing for measurement of more rapid changes in dose rate. This was further demonstrated by measuring the dose delivered by individual photon pulses of a linear accelerator beam. Conclusions: The PMT array system has advantages over CCD camera-based systems for the readout of scintillation light. It provided a more sensitive, more accurate, and faster response to meet the demands of future developments in treatment delivery.

Journal

Medical PhysicsAmerican Association of Physicists in Medicine

Published: Apr 1, 2012

Keywords: brachytherapy; data acquisition; dosimetry; photomultipliers; solid scintillation detectors; photomultiplier; array dosimetry; fiber optic dosimetry; scintillation dosimetry

References

You’re reading a free preview. Subscribe to read the entire article.


DeepDyve is your
personal research library

It’s your single place to instantly
discover and read the research
that matters to you.

Enjoy affordable access to
over 18 million articles from more than
15,000 peer-reviewed journals.

All for just $49/month

Explore the DeepDyve Library

Search

Query the DeepDyve database, plus search all of PubMed and Google Scholar seamlessly

Organize

Save any article or search result from DeepDyve, PubMed, and Google Scholar... all in one place.

Access

Get unlimited, online access to over 18 million full-text articles from more than 15,000 scientific journals.

Your journals are on DeepDyve

Read from thousands of the leading scholarly journals from SpringerNature, Wiley-Blackwell, Oxford University Press and more.

All the latest content is available, no embargo periods.

See the journals in your area

DeepDyve

Freelancer

DeepDyve

Pro

Price

FREE

$49/month
$360/year

Save searches from
Google Scholar,
PubMed

Create folders to
organize your research

Export folders, citations

Read DeepDyve articles

Abstract access only

Unlimited access to over
18 million full-text articles

Print

20 pages / month

PDF Discount

20% off