TY - JOUR AU1 - Kisilevsky, Mila AU2 - Hudson, Christopher AU3 - Flanagan, John G. AU4 - Nrusimhadevara, Ravi Krishna AU5 - Guan, Kit AU6 - Wong, Tien AU7 - Mandelcorn, Mark AU8 - Lam, Wai-Ching AU9 - Devenyi, Robert G. AB - ObjectivesTo estimate the agreement between the macular edema maps (MEMs) of the Retina Module of the Heidelberg Retina Tomograph II (Heidelberg Engineering, Heidelberg, Germany) and contact lens fundus biomicroscopy (FB) and to assess the influence of combining MEM data with the results of short-wavelength automated perimetry (SWAP) and fluorescein angiography (FA) on diagnostic test performance.DesignProspective, observational case series.MethodsTwenty patients (20 eyes) with diabetic retinopathy with or without clinically manifest macular edema (11 and 9 eyes, respectively) were enrolled. All patients underwent full ophthalmologic examination and also MEM assessment, SWAP, and FA.ResultsUsing FB as the “gold standard,” the agreement between the MEMs and FB was very good (Kendall coefficient of concordance, 0.80). Macular edema maps showed good agreement with FA and SWAP (Kendall coefficient, 0.64 and 0.65). Virtually all of the edematous areas detected with MEM but not seen clinically had decreased sensitivity on SWAP and/or fluorescein leakage.ConclusionsMacular edema maps demonstrated very good agreement with FB. Combining the results of FA and SWAP with those of the MEMs provided supporting evidence of concomitant blood-retinal barrier leakage and visual dysfunction, respectively, in areas of early retinal thickening. Prospective studies are ongoing to fully assess the diagnostic test performance of MEMs in the detection of early and progressive diabetic macular edema.Diabetic retinopathy is the leading cause of visual impairment and blindness in the working population of Western countries.Macular edema in particular remains the most common cause of visual impairment and blindness among people with diabetes.The current clinical “gold standard” for the detection of diabetic macular edema (DME) is contact lens stereo fundus biomicroscopy (FB), but subtle changes in retinal thickness are difficult to detect and it is even more challenging to determine change in retinal thickening over time.The first-choice treatment currently available for DME is laser photocoagulation, which is implemented when there is a threat of central vision loss.Conventional laser treatment is effective but may cause complications such as visible laser scars that can enlarge,choroidal neovascularization,and visual field loss.There are 2 possible approaches to reduce potential adverse effects of laser treatment. One is to use less destructive treatment such as subthreshold diode micropulse laser photocoagulation.The other is to improve the accuracy of edema diagnosis, which will allow the start of treatment earlier and the application of photocoagulation in a more precise manner to a smaller area of retina. In fact, the combination of both approaches promises to improve the overall outcome of laser treatment. Furthermore, validation of any new treatment modality in clinical trials will require shorter follow-up periods and smaller sample sizes if objective, sensitive, and reliable techniques for the detection and monitoring of DME are available. Any new technique for the quantitative assessment of DME needs to have a similar, and ideally improved, diagnostic test performance to that of clinical examination.Tong and coworkershave proposed the use of the Heidelberg Retina Tomograph (HRT; Heidelberg Engineering, Heidelberg, Germany) as a screening modality for the detection of sight-threatening diabetic maculopathy and a new scoring system for macular image assessment. The authors reported high specificity and good sensitivity of HRT topographic assessment in a group of diabetic patients with heterogenous retinopathy characteristics.Other studies from our laboratory using the “edema index” methodology have demonstrated good spatial agreement between areas of elevated edema index and clinically detected areas of retinal thickening in diabetic patients.In addition, we have demonstrated good correlation of the HRT edema index with change of visual function, including logMAR visual acuity, conventional automated static perimetry, and short-wavelength automated perimetry (SWAP) in patients undergoing grid laser treatment for clinically significant macular edema.Furthermore, SWAP has been shown to be a sensitive tool for the detection of DME. In many cases, the localized field loss detected by SWAP corresponded to, but was greater than, the area of edema on clinical mapping.The macular edema maps (MEMs) of the Retina Module of the HRT II were subsequently developed. More recently, the HRT II MEMs have been demonstrated to offer improved sensitivity (and comparable specificity) in the detection of DME when compared with the Retinal Thickness Analyzer.The HRT II MEMs have been demonstrated to have good repeatability in patients with diabetic retinopathy and edema.In our experience, however, MEMs tend to detect larger areas of edema than clinical assessment and also on occasion show elevated edema index values in patients without any clinically evident retinal thickening.We hypothesized that the assessment of visual function (ie, SWAP) and blood-retinal barrier integrity (ie, fluorescein angiography [FA]) in cases of controversy between MEMs and clinical assessment may help to confirm or deny the validity of the MEM findings. The aims of the study were to estimate the agreement of the MEMs with contact lens stereo FB and to assess the influence of combining MEM data with the results of SWAP and FA on diagnostic test performance. We compared the results of MEMs with contact lens stereo FB, FA, and SWAP in a group of diabetic patients with and without clinically evident DME.METHODSDiabetic patients were enrolled from the retina clinics at the Toronto Western Hospital, University Health Network, Toronto, Ontario. The study was approved by the research ethics board of the University Health Network, University of Toronto, and the Office of Research Ethics, University of Waterloo, Ontario. The research followed the tenets set out in the Declaration of Helsinki. Informed consent was obtained from all subjects. The sample comprised 20 patients (14 men, 6 women) with a mean ± SD age of 59.15 ± 7.1 years (range, 42-69 years). Nineteen patients had type 2 diabetes, and 1 patient had type 1 diabetes. The mean ± SD duration of diabetes was 15 ± 8.8 years (range, 4-42 years). Inclusion criteria comprised mild-to-moderate diabetic retinopathy, an age range of 40 to 70 years, and a logMAR visual acuity of 0.3 or better. Exclusion criteria comprised lenticular opacities that would impact substantially on fundus visualization, refractive error greater than ±6.00 diopter sphere and/or ±2.50 diopter cylinder, any other eye disorder or disease apart from diabetic retinopathy, previous laser treatment or surgery in the study eye, a family history of glaucoma in a first-degree relative, central nervous system or psychiatric diseases, systemic medications with known central nervous system effects, and unreliable automated perimetry results as defined by the Humphrey Field Analyzer II (Carl Zeiss Meditec Inc, Dublin, Calif). All patients underwent a full ophthalmologic evaluation, including logMAR visual acuity and refraction, slitlamp examination with contact lens stereo FB, FA, SWAP, and HRT II MEM. Pupillary dilation using 1.0% Mydriacyl (Alcon, Fort Worth, Tex) was undertaken for all procedures apart from logMAR visual acuity and refraction. All tests but FA were performed on the same day. One eye of each patient was randomly assigned to the study.MEMs OF THE HRT IIThe HRT II is a confocal laser scanning system for the acquisition and analysis of 3-dimensional images of the posterior segment of the eye. The laser source employed in the HRT II is a diode laser with a wavelength of 670 nm. A 3-dimensional image is generated as a series of 16 to 64 consecutive and equidistant (ie, 16 images every 1 mm of scan depth) 2-dimensional optical section images. Each of the 2-dimensional images consists of 384 × 384 picture elements. The scan field is 15° × 15°. Three sequential tomographic images are automatically scanned on initiation of image acquisition and are averaged to display a mean topography image. The edema index analysis developed by Flanagan and Hudson has been incorporated within the HRT II as part of the Retina Module. The HRT II Retina Module software computes reflectivity, topography, signal width, and edema maps. A full description of the edema index methodology has been detailed previously.The image acquisition was performed by experienced operators and 3 sets (ie, 3 × 3 images) of HRT II images were obtained for each patient.The HRT II MEM images were subsequently assessed by an experienced MEM user. One of the 3 sets of images was chosen based on best image quality. The edema map was divided into 9 sectors using a grid formed by 3 concentric circles centered on the fovea and 4 radial lines extending outward from the innermost of these in the 45°, 135°, 225°, and 315° meridians (9-Zone Analysis). The radius of the innermost circle was 500 μm, and the radii of the second and the third circles were 1000 μm and 1500 μm, respectively. The total field size for the analysis was approximately 10° in diameter (Figure 1). To minimize error in the positioning of the 9-sector grid on the MEM image, the foveal center was located on the basis of clinical and FA assessments and marked on an acetate sheet that detailed each individual's retinal vascular pattern. This also ensured consistency in the positioning of the grid across all imaging modalities (FB, FA, and MEM) (Figure 2). The presence or absence of elevated edema index values was documented for each of the 9 sectors; 180 sectors (ie, 9 sectors × 20 patients) were analyzed for the group as a whole.Figure 1.Nine-sector grid used for the assessment of macular edema map images.Figure 2.This case demonstrates very good agreement between macular edema maps (MEMs) and fundus biomicroscopy (89%), fairly good agreement between MEMs and fluorescein angiography (67%), and good agreement between MEMs and short-wavelength automated perimetry (78%). A, Fundus photograph. B, Fundus biomicroscopy interpretation. C, Fluorescein angiography. D, Macular Edema Module reflectivity map. E, Macular Edema Module edema map. F, Short-wavelength automated perimetry analysis.SHORT-WAVELENGTH AUTOMATED PERIMETRYThe macular visual field of each patient was determined with program 10-2 of the Humphrey Field Analyzer II using short-wavelength stimulus parameters and a standard full-threshold strategy. All subjects had previously undertaken at least 1 SWAP field, thus minimizing learning effects. The SWAP was assessed using a pointwise horizontal and vertical hemifield asymmetry analysis that has been detailed previously,thereby negating the influence of prereceptoral absorption. Abnormality was defined as 7 or more contiguous stimulus locations with statistically significant negative asymmetries. From the total of 68 stimulus locations, only 32 central points corresponding to a 10° field diameter were included in the comparative analysis. These 32 stimulus locations were divided into 9 predefined sectors approximately corresponding to the areas of the MEM (Figure 3). Each predefined sector comprised 4 points (4 points falling on the radial lines were double counted and included into the analysis of both adjacent areas). A given grid sector was considered abnormal if it contained at least 2 statistically significant points with negative SWAP asymmetry values, apart from the central 500-μm radius sector where only 1 abnormal point was required.Figure 3.Grid used for the assessment of short-wavelength automated perimetry results.CONTACT LENS STEREO FBContact lens stereo FB was performed by retinal specialists masked to MEM results. For each patient, all areas of edema were drawn on an acetate sheet that detailed each individual's retinal vascular pattern while the patient was situated at the slitlamp. The 9-sector grid was overlaid on the sheet detailing the areas of retinal thickening to allocate the presence or absence of DME in each sector of the grid.FLUORESCEIN ANGIOGRAPHYFluorescein angiography was performed on all research patients. The interval between FA and the study tests was dictated by routine clinical care of the patient and ranged from the same day to 20 weeks (median, 8 weeks). Late phase and early phase angiograms were used to determine the presence and location of FA leakage according to Early Treatment Diabetic Retinopathy Study Research criteria.All areas of leakage were drawn on an acetate sheet and sectors of involvement were documented in a similar manner to that of FB.STATISTICAL ANALYSISDescriptive statistics such as mean, range, and standard deviation were calculated for case characteristics. Percentage of agreement and the Kendall coefficient of concordance were calculated using the pooled sectors to compare FB results with those of MEMs, SWAP, and FA.RESULTSEleven patients had clinically evident macular edema and 9 patients had no edema by FB. Fourteen patients had an abnormal HRT II MEM. Fourteen patients had localized visual field loss using SWAP. Eleven patients had leakage on FA (Table 1).Table 1. Distribution of Abnormal Test Results*Patient No.FBMEMsFASWAP1XXX2XXXX34XXXX5XX67XXX8XXXX9XXX10XXXX11X1213XXXX14XXX15X16XXXX17X18XXX19XXXX20XX Abbreviations: FA, fluorescein angiography; FB, fundus biomicroscopy; MEMs, macular edema maps; SWAP, short-wavelength automated perimetry. *X denotes abnormal test results; no entry indicates normal results.Across the group, on average 1.85 of 9 sectors were noted as abnormal on FB. Fluorescein angiography, MEMs, and SWAP detected more abnormal sectors than FB (Table 2).Table 2. Number of Areas Defined as AbnormalTestTotal Abnormal Sectors, No.Abnormal Sectors per Patient, Mean, No. (Range)FB37/1801.85/9 (0-5)MEMs56/1802.8/9 (0-9)FA40/1802/9 (0-8)SWAP55/1802.75/9 (0-6) Abbreviations: FA, fluorescein angiography; FB, fundus biomicroscopy; MEMs, macular edema maps; SWAP, short-wavelength automated perimetry.The HRT II MEMs and FB showed very good agreement in identifying areas with DME with a Kendall coefficient of concordance of 0.80 (P<.001). There was substantial agreement between MEMs and FA (Kendall coefficient, 0.64; P<.001) as well as between MEMs and SWAP (Kendall coefficient, 0.65; P<.001) (Table 3). Comparing MEMs with FB as the gold standard, MEMs demonstrated a sensitivity of 92% and specificity of 85%.Table 3. Agreement Between MEMs and FB, FA, and SWAP (in Number of Areas)TestsAgreementTotal Sectors, No.Sectors per Patient, Mean, No.Kendall CoefficientMEMs vs FBPositive34/1801.7/90.80 (P<.001)Negative121/1806.05/9MEMs vs FAPositive24/1801.2/90.64 (P<.001)Negative117/1805.85/9MEMs vs SWAPPositive31/1801.55/90.65 (P<.001)Negative100/1805/9 Abbreviations: FA, fluorescein angiography; FB, fundus biomicroscopy; MEMs, macular edema maps; SWAP, short-wavelength automated perimetry.The mean number of sectors without clinically evident retinal thickening and defined by MEMs as abnormal was 1.1 of 9. Similarly, FA detected a mean of 1.2 additional abnormal sectors while SWAP detected a mean of 1.75 sectors. The mean number of sectors with clinically evident retinal thickening and defined by MEMs as normal was 0.15 of 9, whereas for FA and SWAP, the mean numbers of normal sectors were 1.05 and 0.85, respectively (Table 4).Table 4. Disagreement Between FB, MEMs, FA, and SWAPTestsTotal Sectors, No. (of 180)Sectors per Patient, Mean, No. (of 9)MEMs positive/FB negative22/1431.1/7.15MEMs negative/FB positive3/370.15/1.85FA positive/FB negative24/1431.2/7.15FA negative/FB positive21/371.05/1.85SWAP positive/FB negative35/1431.75/7.15SWAP negative/FB positive17/370.85/1.85 Abbreviations: FA, fluorescein angiography; FB, fundus biomicroscopy; MEMs, macular edema maps; SWAP, short-wavelength automated perimetry.Analyzing only the sectors that were detected as abnormal by MEMs but that did not show retinal thickening on FB, the localized SWAP sensitivity loss was observed in 13 of 22 such sectors and FA leakage in 9 of 22 sectors, and decreased SWAP sensitivity and/or FA leakage was observed in 16 of 22 sectors (Table 5).Table 5. Agreement Between Modalities (MEMs, SWAP, FA) for Additional Abnormal Areas Found Within the MEMsTestTotal Sectors, No.Sectors per Patient, Mean, No.MEMs positive/FB negative22/1431.1/7.15SWAP agreed with MEMs13/220.65/1.1FA agreed with MEMs9/220.45/1.1SWAP and/or FA agreed with MEMs16/220.8/1.1 Abbreviations: FA, fluorescein angiography; FB, fundus biomicroscopy; MEMs, macular edema maps; SWAP, short-wavelength automated perimetry.COMMENTFundus biomicroscopy remains the clinical gold standard for the assessment of DME, but it relies on the subjective interpretation of thickening of an essentially transparent retina, has large interobserver variability, and lacks objective quantification. The development of a reproducible, objective technique to identify retinal edema and to monitor change in edema is widely recognized to be essential for the improved clinical management of DME per se and for the evaluation of new therapeutic protocols.Currently available imaging technologies to assess DME include optical coherence tomography (OCT), the Retinal Thickness Analyzer, and the Macular Edema Module of the HRT II. Optical coherence tomography is widely used for the assessment of macular disease with different etiologies. Optical coherence tomography has very good agreement with contact lens FB,although it has been reported to be prone to artifacts arising in the process of macular surface map reconstruction.The Retinal Thickness Analyzer works on the principle of slitlamp FB and in our experience shows poor sensitivity and moderate specificity in the assessment of the retinal thickness of diabetic patients.The Macular Edema Module represents a relatively new and promising application for the detection and follow-up of DME. The Macular Edema Module generates an edema index that reflects increase in retinal thickness and reduction in retinal reflectivity due to changes in retinal refractive index in areas of edema. The reliance of the Macular Edema Module technique on 2 optical principles for the detection of edema is thought to increase the sensitivity.Compared to OCT, the HRT II has inferior axial resolution; however, it has far superior spatial resolution and thereby avoids the undersampling in the periphery of the image that is inherent in the OCT macular maps. This study and previous work from our laboratorydemonstrated that MEMs offer very good sensitivity and similar specificity in detecting macular edema compared to FB. The Macular Edema Module generates high spatial resolution topographic edema maps and, unlike any of the current imaging technologies, offers facility for the precise alignment of images acquired over successive visits, an essential feature for any sequential analysis. The extent to which all imaging technologies are influenced by the ocular media, corneal drying, and excessive eye movements is uncertain, but experience suggests that such factors impact the diagnostic test performance of every modality.The lower specificity of MEMs when compared with FB may be due to the fact that the Macular Edema Module detects edema-related changes that are not yet clinically manifest using FB. The failure of FB to correlate with alternative objective techniques for the assessment of retinal thickness has been reported by previous investigators.Brown et alproposed the term subclinical edemafor such cases. The assessment of the blood-retinal barrier integrity and short-wavelength pathway function in areas of possible subclinical DME was used in our study to validate the MEM results. The presence of edema suggests blood-retinal barrier breakdown, and areas of retinal edema largely, although not completely, coincide with areas of fluorescein leakage. Similarly, localized SWAP sensitivity loss has been shown to correspond with the clinical mapping of the area of DME, but the extent of this loss was found to be generally greater than that suggested by clinical assessment.Although SWAP visual field defects in diabetic maculopathy may also be caused by ischemic damage in the absence of DME,the larger area of defect most likely reflects subtle changes in retinal thickness, particularly given the clinical characteristics of the patients described in this study.We assessed the MEMs using a 9-sector grid rather than define the image as normal or abnormal as a whole. This approach allowed more precise assessment of MEMs and a topographic comparison of different techniques used for the detection of abnormality associated with the occurrence of DME. Macular edema maps showed excellent agreement with FB but tended to show a greater extent of abnormality. Almost all sectors identified as abnormal using MEMs but normal with FB and revealed some abnormality on SWAP and/or FA, suggesting the presence of functional and structural changes. Thus, combining the findings of FA and SWAP with the MEM scans provided supporting evidence of abnormality in areas of subclinical retinal edema. Ideally, a prospective study with a larger group of patients will determine whether these positive MEM findings will progress to clinically evident retinal edema. Such a study, also incorporating OCT, is currently in progress in our laboratory.Correspondence:Christopher Hudson, PhD, University of Toronto, Toronto Western Hospital, Main Pavilion 6-206, 399 Bathurst St, Toronto, Ontario, Canada M5T 2S8 (chudson@sciborg.uwaterloo.ca).Submitted for Publication:October 5, 2004; accepted March 14, 2005.Financial Disclosure:Dr Flanagan has received payment as a consultant and has received research funding from Heidelberg Engineering, Heidelberg, Germany.Funding/Support:This work was funded by the Canadian Institutes of Health Research, Ottawa, Ontario (New Investigator Award, C.H.; Operating Grant, C.H. and J.G.F.); and by a Premier's Research Excellence Award (C.H.).REFERENCESRKleinBEKleinSEMossKJCruickshanksThe Wisconsin epidemiology study of diabetic retinopathy, XVII: the 14 year incidence and progression of diabetic retinopathy and associated risk factors in type I diabetes.Ophthalmology1998105180118159787346RKleinBEKleinSEMossMDDavisDLDeMetsThe Wisconsin epidemiologic study of diabetic retinopathy, IV: diabetic macular edema.Ophthalmology198491146414746521986RKleinSEMossBEKleinMDDavisDLDeMetsThe Wisconsin epidemiologic study of diabetic retinopathy, IX: the incidence of macular edema.Ophthalmology198996150115102587045CHudsonJGFlanaganDMcLeodA clinical vision science perspective of the management of diabetic macular oedema: focus on diabetic retinopathy.Excerpta Medica199964144Early Treatment Diabetic Retinopathy Study Research GroupEarly photocoagulation for diabetic retinopathy: ETDRS report number 9.Ophthalmology199198(suppl 5)7667852062512Early Treatment Diabetic Retinopathy Study Research GroupFocal photocoagulation treatment of diabetic macular edema. Relationship of treatment effect to fluorescein angiographic and other retinal characteristics at baseline: ETDRS report No. 19.Arch Ophthalmol1995113114411557661748HLewisAPSchachatMHHaimannChoroidal neovascularization after laser photocoagulation for diabetic macular edema.Ophthalmology1990975035101691477CHudsonJGFlanaganGSTurnerHCChenLBYoungDMcLeodInfluence of laser photocoagulation for clinically significant diabetic macular oedema (DMO) on short-wavelength and conventional automated perimetry.Diabetologia199841128312929833934JKLuttrullDCMuschMAMainsterSubthreshold diode micropulse photocoagulation for the treatment of clinically significant diabetic macular oedema.Br J Ophthalmol200589748015615751LTongAAngSAVernonSensitivity and specificity of a new scoring system for diabetic macular oedema detection using a confocal laser imaging system.Br J Ophthalmol200185343911133709CHudsonJGFlanaganGSTurnerDMcLeodScanning laser tomography Z-profile signal width as an objective index of retinal thickening due to macular oedema.Br J Ophthalmol1998821211309613376CHudsonJGFlanaganGSTurnerHCChenLYoungDMcLeodCorrelation of a scanning laser derived oedema index and visual function following grid laser treatment for diabetic macular oedema.Br J Ophthalmol20038745546112642310CHudsonJGFlanaganGSTurnerHCChenLBYoungDMcLeodShort-wavelength sensitive visual field loss in patients with clinically significant diabetic macular oedema.Diabetologia1998419189289726594KGuanCHudsonJGFlanaganComparison of Heidelberg Retina Tomograph II and Retinal Thickness Analyzer in the assessment of DME.Invest Ophthalmol Vis Sci20044561061614744905Early Treatment Diabetic Retinopathy Study Research GroupClassification of diabetic retinopathy from fluorescein angiograms: ETDRS report number 11.Ophthalmology1991988078222062514GHBresnickDiabetic macular edema.Ophthalmology1986939899973531959JCBrownSDSolomonSBBresslerAPSchachatCDiBernardoNMBresslerDetection of diabetic foveal edema: contact lens biomicroscopy compared with optical coherence tomography.Arch Ophthalmol200412233033515006844RRaySSStinnettGJJaffeEvaluation of image artifact produced by optical coherence tomography of retinal pathology.Am J Ophthalmol2005139182915652824MShahidiYOguraNBlairMMRusinRZeimerRetinal thickness analysis for quantitative assessment of diabetic macular edema.Arch Ophthalmol1991109111511191867554ARemkyOArendSHendricksShort-wavelength automated perimetry and capillary density in early diabetic maculopathy.Invest Ophthalmol Vis Sci20004127428210634631 TI - Agreement of the Heidelberg Retina Tomograph II Macula Edema Module With Fundus Biomicroscopy in Diabetic Maculopathy JF - JAMA Ophthalmology DO - 10.1001/archopht.124.3.337 DA - 2006-03-01 UR - https://www.deepdyve.com/lp/american-medical-association/agreement-of-the-heidelberg-retina-tomograph-ii-macula-edema-module-xdiFrM2L0W SP - 337 EP - 342 VL - 124 IS - 3 DP - DeepDyve ER -