TY - JOUR AU1 - G, Pronost, AU2 - M, Ikeda, AU3 - T, Nakamura, AU4 - H, Sekiya, AU5 - S, Tasaka, AB - Abstract Radioactivity from radon is a major threat for high-precision low-energy physics experiments like the ones in the Kamioka mine. We developed a new high-sensitivity radon monitoring system and conducted systematic radon concentration measurements for the first time in Kamioka. The system consists of portable radon detectors with a capacity of 1 L and new electronics based on a Raspberry Pi [Raspberry Pi documentation. (Available at: https://www.raspberrypi.org/documentation/)]. These radon detectors measure the radon in the air with electrostatic collection and a PIN photo-diode. We measured the absolute humidity dependence of the 1-L radon detector for air as $$C_{F} (A_{H}) = (12.86 \pm 0.40) - (1.66 \pm 0.19) \sqrt{A_{H}}$$$$\mathrm{(counts/day)/(Bq/m^3)}$$. The background level of the 1-L radon detector is $$0.65\pm0.15$$ (stat.) counts/day. This corresponds to a detection limit of $$\sim0.4$$ Bq/m$$^3$$ in a one-day measurement. Data were collected for a period of more than one year with twenty-one 1-L radon detectors in the Kamioka mine. They indicate seasonal and day–night variations in radon concentration within the mine. These results also allow us to confirm the stability of the new Raspberry Pi electronics. 1. Introduction $$^{222}$$Rn (called radon hereafter) is a radioactive gas abundant in most underground places. It is produced in the 4n+2 decay chain of $$^{238}$$U, the so-called “radium series”. Its decay chain is as follows: $$^{222}_{\,86}\mathrm{Rn} \xrightarrow[3.82\,\mathrm{d}] {\alpha_1} \, ^{218}_{\,84}\mathrm{Po} \xrightarrow[3.1\,\mathrm{min}] {\alpha_2} \, ^{214}_{\,82}\mathrm{Pb} \xrightarrow[26.8\,\mathrm{min}] {\beta^-} \, ^{214}_{\,83}\mathrm{Bi} \xrightarrow[19.9\,\mathrm{min}] {\beta^-} \, ^{214}_{\,84}\mathrm{Po} \xrightarrow[164.3\,\mu \mathrm{s}]{\alpha_3} \, ^{210}_{\,82}\mathrm{Pb}$$$$\xrightarrow[22.26 \, \mathrm{yr}] {\beta^-} \, ^{210}_{\,83}\mathrm{Bi} \xrightarrow[5.01 \, d] {\beta^-\ }\, ^{210}_{\,84}\mathrm{Po} \xrightarrow[138.38\,\mathrm{d}] {\alpha_4} \, ^{206}_{\,82}\mathrm{Pb}$$, where $$\alpha_{1}$$, $$\alpha_{2}$$, $$\alpha_{3}$$, and $$\alpha_{4}$$ are the different $$\alpha$$ particles in this decay chain with energies of $$E(\alpha_1)=5.590$$ MeV, $$E(\alpha_2)= 6.002$$ MeV, $$E(\alpha_3)=7.833$$ MeV, and $$E(\alpha_4)=5.407$$ MeV [2]. Due to the production of $$\alpha$$ and $$\beta^-$$ particles over a long period, $$^{222}$$Rn is a serious source of background for low-energy physics experiments and underground experiments in particular. In addition, as a radioactive gas, workers in underground environments can be exposed to high concentrations of radon, which can lead to serious health issues [3]. These two issues suggest the need to monitor the radon concentration in underground environments. The Kamioka mine in the Gifu Prefecture of Japan hosts several high-precision, low-energy experiments, including Super-Kamiokande [4] , KamLAND [5] , XMASS [6] , CANDLES [7] , and NEWAGE [8] . To monitor the radon concentration in the air of the mine, we developed a precise and cost-effective system using a 1-L detector with Raspberry Pi-based electronics. 2. Detector design To measure the radon concentration in the air, we used the electrostatic collection method of the daughter nuclei of $$^{222}$$Rn, combined with the energy measurement of alpha particles from the daughter-nuclei decays. This energy measurement is performed using a PIN photo-diode. Previous publications [9] showed that $$90\%$$ of the isotopes produced in the $$^{222}$$Rn-nuclei decay chain are positive ions. Therefore, they can be collected on a PIN photo-diode on which a negative high voltage (HV) is applied. $$\alpha_{2}$$, $$\alpha_{3}$$, and $$\alpha_{4}$$, referring to the $$\alpha$$ particles from the $$^{218}$$Po, $$^{214}$$Po, and $$^{210}$$Po decays, respectively, can be detected by the PIN photo-diode if the polonium nucleus has been collected. Since $$^{222}$$Rn nuclei are not ions, they are unlikely to be collected on the PIN photo-diode; $$\alpha_{1}$$ are then unlikely to be detected. Only $$\alpha_{2}$$, $$\alpha_{3}$$, and $$\alpha_{4}$$ measurements are performed. Figure 1 shows a schematic of the new 1-L radon detectors that we use in the Kamioka mine; this schematic is of the same type as that shown in Fig. 4 of Ref. [10]. These detectors are made of stainless-steel cases, open at the bottom in order to allow air from the atmosphere to enter the detector. A Hamamatsu S3590-09 PIN photo-diode is located at the top of the detector and is connected via a feed-through to an HV supplier, a preamplifier, and a shaping amplifier. The air opening at the bottom of the detector is covered by three different layers. From inside to outside, these are a stainless-steel mesh, a microporous membrane, and a black cover. The stainless-steel mesh is used as an electric ground in continuity with the detector case. It consists of 100 mesh per inch. The microporous membrane is used as a filter to remove dust from the incoming air and thus the daughter nuclei of the radon. This allows us to collect only polonium, bismuth, and lead positive ions from radon decays occurring within the detector case. The microporous membrane used is a polypropylene membrane produced by the CELGARDTM company (product’s name: CELGARD© 2400). The last cover is a piece of black fabric, which is used to shield the inside of the detector, and thus the PIN photo-diode, from external light. Fig. 1. View largeDownload slide Schematic of a 1-L radon detector. Measurements are in mm. Fig. 1. View largeDownload slide Schematic of a 1-L radon detector. Measurements are in mm. The ventilation rate of the 1-L radon detector has been calculated to be about seven times per hour with natural ventilation. The computation was done using the theoretical equation and experimental values from Ref. [11]. This large value is owing to the 80-mm-diameter ventilation window. 3. Electronics The HV supplier and the amplifier circuit are described in Fig. 2, and a picture is shown in Fig. 3. An HV of $$-$$120 V is supplied to the whole detector. This bias voltage allows positive ions from the radon decays to be collected on the PIN photo-diode. The value of the HV was determined as the voltage for which the collection of $$^{214}$$Po and $$^{218}$$Po positive ions is at its maximum. Fig. 2. View largeDownload slide Circuit scheme of the HV supplier and amplifier circuit. The combination of resistors R12 and R13 leads to a $$7.5$$ gain. The $$\pm12$$ V input signal is connected to CN1 (1,5,6); the positive output signal is connected to CN1 (2,3); the HV is connected to CN2 (7,8). Connectors are set between JP1 (2,3), between JP1 (4,5), and between JP2 (2,3). Fig. 2. View largeDownload slide Circuit scheme of the HV supplier and amplifier circuit. The combination of resistors R12 and R13 leads to a $$7.5$$ gain. The $$\pm12$$ V input signal is connected to CN1 (1,5,6); the positive output signal is connected to CN1 (2,3); the HV is connected to CN2 (7,8). Connectors are set between JP1 (2,3), between JP1 (4,5), and between JP2 (2,3). Fig. 3. View largeDownload slide Picture of the amplifier circuit. View from above. Dimensions: 2 $$\times$$ 5 cm. Fig. 3. View largeDownload slide Picture of the amplifier circuit. View from above. Dimensions: 2 $$\times$$ 5 cm. The charge from the PIN photo-diode is converted to voltage pulses by the preamplifier circuit. The pulse is then shaped using the operational amplifier AD847JRZ [Ref: Analog Devices AD847 datasheet http://www.analog.com/media/en/technical-documentation/data-sheets/AD847.pdf], inverted, and amplified. The amplification gain of the amplifier circuit is $$7.5$$; this can be reduced by adding a resistor at RR1 or increased by removing either R12 or R13. The signal is then sent to an analog-to-digital-converter (ADC) board specially developed to be used with Raspberry Pi systems. In order to measure the equivalent-noise-charge (ENC) from the preamplifier, the PIN photo-diode has been irradiated with $$\gamma$$-rays from an $$^{241}$$Am source. The 59.4 keV photoelectric peak was measured, and allowed measurement of the value of the ENC as 530 electrons (FWHM). The ADC board circuit is described in Fig. 4. It consists of a comparator circuit, a peak-holder circuit, an 8-bit ADC (MAX153CWP) [12], a pulse timing circuit, and a latch circuit. It accepts an input signal via a 50 $$\Omega$$-impedance LEMO connector with a maximal amplitude of 1000 mV. This amplitude is converted into a channel number by the peak-holder circuit and the 8-bit ADC, and then sent to the Raspberry Pi. Fig. 4. View largeDownload slide Circuit scheme of the ADC board used for the Raspberry Pi B, B+, 2B, and 3B systems. For Raspberry Pi B+, 2B, and 3B, the circuit also contains the circuit shown in Fig. 5. Fig. 4. View largeDownload slide Circuit scheme of the ADC board used for the Raspberry Pi B, B+, 2B, and 3B systems. For Raspberry Pi B+, 2B, and 3B, the circuit also contains the circuit shown in Fig. 5. Fig. 5. View largeDownload slide Additional part of the circuit scheme of the ADC board used for the Raspberry Pi B+, 2B, and 3B systems. Fig. 5. View largeDownload slide Additional part of the circuit scheme of the ADC board used for the Raspberry Pi B+, 2B, and 3B systems. Raspberry Pi B, B+, 2B, and 3B [1] are currently used in the Kamioka mine. Due to the architectural differences between the different types of Raspberry Pi systems, two different versions of the ADC board have been developed. The first version, called ADC-B, is shown in Fig. 6. It is used with Raspberry Pi B, to which it is connected via a 26-GPIO-pins (General Purpose Input/Output) connector. The second version, called ADC-B+, is shown in Fig. 7. It is used with Raspberry Pi B+, 2B, and 3B and is connected to the Raspberry Pi via a 40-GPIO-pin connector. The circuit of the ADC-B+ is similar to that of the ADC-B; however, because of the additional pins, four additional analog inputs are added to the circuit. These inputs accept signals with a maximal amplitude of 5000 mV and process them by a 4-channel 10-bit ADC (ADC104S021CIMM) [13]. These additional inputs are used to record environmental data (temperature, humidity, flow rate, etc.). Fig. 6. View largeDownload slide Picture of the ADC board version B, used for Raspberry Pi B systems. View from above. Dimensions: 5.5 $$\times$$ 8.5 cm. Fig. 6. View largeDownload slide Picture of the ADC board version B, used for Raspberry Pi B systems. View from above. Dimensions: 5.5 $$\times$$ 8.5 cm. Fig. 7. View largeDownload slide Picture of the ADC board version B+, used for Raspberry Pi B+, 2B, and 3B systems. View from above. Dimensions: 5.5 $$\times$$ 8.5 cm. Fig. 7. View largeDownload slide Picture of the ADC board version B+, used for Raspberry Pi B+, 2B, and 3B systems. View from above. Dimensions: 5.5 $$\times$$ 8.5 cm. The main part of the circuit is common to both versions and is described in Fig. 4; the additional circuit for the four additional inputs is described in Fig. 5. These boards are manufactured by WIN Electronics Co., Ltd.1 3.1. Data acquisition The ADC board is monitored by software to perform data acquisition. This software was developed using C++ and uses the wiringPi framework [14] to communicate with the ADC board through the GPIO-pin connector. This software was developed to record data with fixed time interval measurements. However, the CPU speed on the Raspberry Pi systems limits the minimal interval that can be used. To estimate this minimal time interval, hardware-limitation tests have been performed using a Raspberry Pi 3B equipped with a 48-MB/s TOSHIBA 32-GB micro-SD card. Table 1 summarizes the number of times that the actual interval was different than the expected interval. Table 1. Summary of the study of the minimal time interval measurement. The “Different intervals” columns show the ratio (third column) and number of times per day (fourth column) that the actual interval was not the same as the expected interval. Only statistical uncertainties are shown.   Expected interval Actual mean interval Different intervals $$\%$$ $$\#$$ per day $$1$$ sec $$1.59 \pm 0.50$$ sec $$59.6201 \pm 1.1182 \%$$ $$51\,806 \pm 967$$ $$2$$ sec $$2.01 \pm 0.21$$ sec $$0.2717 \pm 0.0008 \%$$ $$117 \pm 1$$ $$3$$ sec $$3.01 \pm 0.15$$ sec $$0.1292 \pm 0.0005 \%$$ $$37 \pm 1$$ $$4$$ sec $$4.00 \pm 0.13$$ sec $$0.0356 \pm 0.0002 \%$$ $$8 \pm 0$$ ($$0.05$$) $$5$$ sec $$5.00 \pm 0.10$$ sec $$0.0296 \pm 0.0002 \%$$ $$5 \pm 0$$ ($$0.02$$)   Expected interval Actual mean interval Different intervals $$\%$$ $$\#$$ per day $$1$$ sec $$1.59 \pm 0.50$$ sec $$59.6201 \pm 1.1182 \%$$ $$51\,806 \pm 967$$ $$2$$ sec $$2.01 \pm 0.21$$ sec $$0.2717 \pm 0.0008 \%$$ $$117 \pm 1$$ $$3$$ sec $$3.01 \pm 0.15$$ sec $$0.1292 \pm 0.0005 \%$$ $$37 \pm 1$$ $$4$$ sec $$4.00 \pm 0.13$$ sec $$0.0356 \pm 0.0002 \%$$ $$8 \pm 0$$ ($$0.05$$) $$5$$ sec $$5.00 \pm 0.10$$ sec $$0.0296 \pm 0.0002 \%$$ $$5 \pm 0$$ ($$0.02$$) Table 1. Summary of the study of the minimal time interval measurement. The “Different intervals” columns show the ratio (third column) and number of times per day (fourth column) that the actual interval was not the same as the expected interval. Only statistical uncertainties are shown.   Expected interval Actual mean interval Different intervals $$\%$$ $$\#$$ per day $$1$$ sec $$1.59 \pm 0.50$$ sec $$59.6201 \pm 1.1182 \%$$ $$51\,806 \pm 967$$ $$2$$ sec $$2.01 \pm 0.21$$ sec $$0.2717 \pm 0.0008 \%$$ $$117 \pm 1$$ $$3$$ sec $$3.01 \pm 0.15$$ sec $$0.1292 \pm 0.0005 \%$$ $$37 \pm 1$$ $$4$$ sec $$4.00 \pm 0.13$$ sec $$0.0356 \pm 0.0002 \%$$ $$8 \pm 0$$ ($$0.05$$) $$5$$ sec $$5.00 \pm 0.10$$ sec $$0.0296 \pm 0.0002 \%$$ $$5 \pm 0$$ ($$0.02$$)   Expected interval Actual mean interval Different intervals $$\%$$ $$\#$$ per day $$1$$ sec $$1.59 \pm 0.50$$ sec $$59.6201 \pm 1.1182 \%$$ $$51\,806 \pm 967$$ $$2$$ sec $$2.01 \pm 0.21$$ sec $$0.2717 \pm 0.0008 \%$$ $$117 \pm 1$$ $$3$$ sec $$3.01 \pm 0.15$$ sec $$0.1292 \pm 0.0005 \%$$ $$37 \pm 1$$ $$4$$ sec $$4.00 \pm 0.13$$ sec $$0.0356 \pm 0.0002 \%$$ $$8 \pm 0$$ ($$0.05$$) $$5$$ sec $$5.00 \pm 0.10$$ sec $$0.0296 \pm 0.0002 \%$$ $$5 \pm 0$$ ($$0.02$$) From these results, we can see that the CPU speed of the Raspberry Pi cannot perform 1 measurement per second as this configuration leads to obtaining fewer data than when asking for 1 measurement every 2 seconds. Hence, we determined that the minimal interval that can be used without critical data loss is 2 seconds. Additional tests showed that electronics are able to support without loss an input of $$\sim23$$ kHz in the case of Raspberry Pi B and B+, and of $$\sim28$$ kHz in the case of Raspberry Pi 2B and 3B. At this rate, the current version of the software needs to be run with a 4-sec (or less) interval in order to avoid data loss. Taking into account the calibration factor shown later, and assuming a humidity of $$10$$$$\mathrm{g/m}^{3}$$, this rate theoretically corresponds to a maximal radon concentration of $$\sim260$$$$\mathrm{MBq/m}^{3}$$ in the case of Raspberry Pi B and B+, and of $$\sim315$$$$\mathrm{MBq/m}^{3}$$ in the case of Raspberry Pi 2B and 3B. 4. Calibration of the radon detector Previous studies showed that the $$^{222}$$Rn daughter nuclei can be collected and neutralized by water molecules in the air [16]. Therefore, the detection efficiency depends on the absolute humidity in the environment. A study of the humidity dependence of the calibration factor $$C_{F}$$ (in $$\mathrm{(counts/day)/(Bq/m^3)}$$) was thus performed. The calibration system was a closed loop including a 1-L detector, a humidity meter, a radon source, and an air pump. A refrigerator was used to control the dew-point temperature of the system. A schematic of the calibration system is shown in Fig. 8. The calibration was performed over two different periods. First, we measured the high-humidity dependence ($$>6$$$$\mathrm{g/m^3}$$); then, we measured the low-humidity dependence ($$<1$$$$\mathrm{g/m^3}$$). The humidity was measured with a MICHELL sensor (PCMini52) during the first campaign and with a VAISALA dew-point meter (DMT340) during the second. The MICHELL sensor measured the relative humidity and air temperature, whereas the VAISALA sensor measured the dew point and the air temperature. The conversion to absolute humidity was performed using the VAISALA humidity formula in the $$[-20,50]^{\circ}$$C approximation [17]. The actual radon concentration was measured with a Pylon scintillator counter (Lucas cell 300A) [18]. The accuracy of the Pylon scintillator counter has been estimated to be $$\pm4\%$$; this value is used as a systematic error in the calibration. Fig. 8. View largeDownload slide Schematic of the 1-L Rn detector calibration system. “Dew” indicates the position of the humidity sensor, “MF” indicates the position of the mass flow controller used to monitor the air flow in the loop, “P” indicates the position of the air pump, “Ref” indicates the position of the sampling point used to measure the actual radon concentration, and “$$^{226}$$Ra” indicates the position of the $$^{226}$$Ra source. The arrows show the direction of the air flow. Fig. 8. View largeDownload slide Schematic of the 1-L Rn detector calibration system. “Dew” indicates the position of the humidity sensor, “MF” indicates the position of the mass flow controller used to monitor the air flow in the loop, “P” indicates the position of the air pump, “Ref” indicates the position of the sampling point used to measure the actual radon concentration, and “$$^{226}$$Ra” indicates the position of the $$^{226}$$Ra source. The arrows show the direction of the air flow. The peak from $$^{218}$$Po $$\alpha$$ decay can be overlapped with the peak from $$^{216}$$Po $$\alpha$$ decay in the Th-series decay chains. Hence, the number of counts used to compute the radon concentration is the integral of the peak from the $$^{214}$$Po $$\alpha$$ decays. This approach was also used in Ref. [19]. Figure 9 shows the measured calibration factors in $$\mathrm{(counts/day)/(Bq/m^3)}$$ as a function of the absolute humidity. The data were found to be best described by $$f(x) = p0 - p1 \times \sqrt{x}$$, as in Ref. [19]. The fit result is shown as: Fig. 9. View largeDownload slide 1-L detector-calibration factor in $$\mathrm{(counts/day)/(Bq/m^3)}$$ as a function of the absolute humidity. The uncertainties of the absolute humidity for the two last points are $$0.06$$$$\mathrm{g/m^3}$$ and $$0.12$$$$\mathrm{g/m^3}$$ (from left to right) and are therefore not visible here. The red solid line shows the fit with a function $$f(A_{H}) = p0 - p1 \times \sqrt{A_{H}}$$, with $$A_{H}$$ being the absolute humidity and $$p0$$ and $$p1$$ being the function’s parameters. The white markers refer to the data points from Ref. [10]. These data points were not used to determine the fit function. Fig. 9. View largeDownload slide 1-L detector-calibration factor in $$\mathrm{(counts/day)/(Bq/m^3)}$$ as a function of the absolute humidity. The uncertainties of the absolute humidity for the two last points are $$0.06$$$$\mathrm{g/m^3}$$ and $$0.12$$$$\mathrm{g/m^3}$$ (from left to right) and are therefore not visible here. The red solid line shows the fit with a function $$f(A_{H}) = p0 - p1 \times \sqrt{A_{H}}$$, with $$A_{H}$$ being the absolute humidity and $$p0$$ and $$p1$$ being the function’s parameters. The white markers refer to the data points from Ref. [10]. These data points were not used to determine the fit function. \begin{equation} C_{F} (A_{H}) = (12.86 \pm 0.40) - (1.66 \pm 0.19) \sqrt{A_{H}}, \end{equation} (1) with $$C_{F}$$ being the calibration factor in $$\mathrm{(counts/day)/(Bq/m^3)}$$ and $$A_{H}$$ being the absolute humidity in $$\mathrm{g/m^3}$$. The uncertainties include both statistical and systematic errors. This fit result is consistent with the previous study performed in Ref. [10], as is shown in Fig. 9 with the data points from Fig. 6 in Ref. [10]. In order to measure the background level of the 1-L detector, we sealed a 1-L detector in a stainless-steel container purged with pure air. The background level was measured over one month as $$0.65\pm0.15$$$$\alpha_{3}$$ counts/day. This allowed us to determine the detection lower limit with Currie’s method [20] as $$\sim0.4$$$$\mathrm{Bq/m^{3}}$$ for a $$10$$$$\mathrm{g/m^{3}}$$ humidity, for one day of measurement. Due to the differences in the HV feed-through manual manufacturing and in the PIN photo-diode settings, the detector response is slightly different for each detector. In order to take this into account, each detector was individually calibrated. This calibration was done in a high-radon-concentration environment ($$>500$$$$\mathrm{Bq/m^{3}}$$ in the mine) where each detector was deployed together with a reference detector (ionization chamber detector, SAPHYMO AlphaGUARD PQ2000) for several days. The stability of this calibration was checked with the deployment of a “reference” detector over the full calibration period. The relative deviation between the AlphaGUARD detector and a 1-L detector is used to determine a correction factor, which is applied to the detector result. The correction factors used for the detectors presented in this paper are shown in Fig. 10. The error of the measured correction factor is dominated by the statistics due to the relatively short measurement period for quick deployment of $$\sim10$$ detectors. Fig. 10. View largeDownload slide Correction factor for the detectors used in Sect. 6. The $$x$$-axis shows the detector number. The correction factor for the reference detector is shown in red (detector #0). Fig. 10. View largeDownload slide Correction factor for the detectors used in Sect. 6. The $$x$$-axis shows the detector number. The correction factor for the reference detector is shown in red (detector #0). 5. Data analysis A framework, based on C++ and on ROOT/CERN [21], was developed to automatically analyze the data from many different radon detectors in the Kamioka mine. Using a standard approach from Ref. [22], the spectrum obtained from the ADC board is analyzed. Cuts are defined to integrate the spectrum between four intervals, three of which correspond to the main $$\alpha$$ decays from the $$^{222}$$Rn decay chain ($$\alpha_2$$, $$\alpha_3$$, and $$\alpha_4$$) and the last to the $$\alpha$$ decay of $$^{212}$$Po ($$E(\alpha_\mathrm{Th}) =8.784$$ MeV) in the $$^{232}$$Th-series decay chains. The measurement of the number of counts from $$^{212}$$Po $$\alpha$$ decays allows estimation of the contamination from $$^{216}$$Po $$\alpha$$ decays ($$E = 6.788$$ MeV) in the $$^{218}$$Po $$\alpha$$-decay peak, as explained in Sect. 4. $$\alpha$$ particles from $$^{212}$$Po are denoted by $$\alpha_\mathrm{Th}$$ in the following. The cuts are defined separately for each detector because small variations of the gain between each amplifier can lead to variations in the peak positions. Figure 11 shows an example of the cuts performed for one detector. Fig. 11. View largeDownload slide Spectrum from one 1-L radon detector. The red lines show the cuts used to determine the peak positions. This detector is using a brand-new PIN photo-diode (less than one month exposure), hence the absence of a $$^{210}$$Pb peak ($$\alpha_{4}$$). Fig. 11. View largeDownload slide Spectrum from one 1-L radon detector. The red lines show the cuts used to determine the peak positions. This detector is using a brand-new PIN photo-diode (less than one month exposure), hence the absence of a $$^{210}$$Pb peak ($$\alpha_{4}$$). The main difficulties with the automatic analysis of dozens of detectors is the possible fluctuation of the peak positions. Each time a detector HV is restarted or the PIN photo-diode or amplifier is replaced, the gain can vary, leading to a change in the peak position. The software is thus able to determine whether the peak positions change and raises an error if this happens. Human action is then needed to determine the new peak positions. In normal operation, this maintenance is not needed for more than two years. It is mainly needed after power failure in the mine, and if the PIN photo-diode or the amplifier is replaced. Data analyses are performed on a server in the Kamioka network. However, thanks to their higher performances, data analyses can also be performed directly on the Raspberry Pi 2B and 3B systems. This allows the use of these electronics as standalone units outside of the Kamioka network. 6. Results Twenty-one 1-L detectors using Raspberry Pi electronics are monitoring the radon concentration in the air at various positions in the Kamioka mine. In this section, the measurements from a selection of these detectors are shown and discussed. Figure 12 shows the positions of these 21 detectors in the Kamioka mine, and Fig. 13 zooms in on the Super-Kamiokande area. Fig. 12. View largeDownload slide Map of the experimental area of the Kamioka mine. The red points refer to the detectors’ locations. The numbers indicate the devices used in this article. The blue point refers to the anemometer’s position. Fig. 12. View largeDownload slide Map of the experimental area of the Kamioka mine. The red points refer to the detectors’ locations. The numbers indicate the devices used in this article. The blue point refers to the anemometer’s position. Fig. 13. View largeDownload slide Zoom in on the Super-Kamiokande experimental area. The red points indicate the detectors’ locations. The numbers are an indication for the detectors used in this article, with number 1 showing a reference from Fig. 12. The blue lines indicate the airtight doors’ positions. Fig. 13. View largeDownload slide Zoom in on the Super-Kamiokande experimental area. The red points indicate the detectors’ locations. The numbers are an indication for the detectors used in this article, with number 1 showing a reference from Fig. 12. The blue lines indicate the airtight doors’ positions. 6.1. Radon concentration in the mine Figure 14 shows the radon concentrations from January 2016 to October 2017 in the tunnels of the mine from the detectors #1 and #2 in Fig. 12 (directly exposed to the air of the mine). The humidity in these areas ranges from 8–12 g/m$$^{3}$$. A clear seasonal variation in the radon concentration is observed. In the approximate period between April to November, the radon concentration is high ($$>1000$$ Bq/m$$^3$$), whereas it is low ($$<500$$ Bq/m$$^3$$) between November and April. In the Atotsu tunnel, the main access tunnel to the Super-Kamiokande area, we measure the speed and direction of the wind with an anemometer, whose position is indicated in Fig. 12. A correlation is observed between the radon concentration in the Atotsu tunnel and the direction of the wind, as shown in Fig. 15. The radon concentration in the Atotsu tunnel is high when the wind comes from inside the mine and low when it comes from outside. Our interpretation is that fresh air with low radon concentration is brought into the mine when wind is coming in from outside, whereas the high-radon-concentration air is pushed out of the mine when the wind is blowing in the opposite direction. An increase of the radon concentration is observed from 10 January 2017. This increase is due to a temporary modification of the water flow in the mine, causing radon-rich water to be near the detector. Fig. 14. View largeDownload slide Radon concentration (2-day-averaged) measured in Kamioka’s mine areas without fresh-air input. The dotted lines indicate the 148-Bq/m$$^3$$ US limit for indoor radon concentration [3]. The top figure (Atotsu tunnel) starts around 7 April 2016 as the detector was not deployed in the mine beforehand. The error bars include statistical uncertainties, as well as the uncertainty on the detector’s correction factor. Fig. 14. View largeDownload slide Radon concentration (2-day-averaged) measured in Kamioka’s mine areas without fresh-air input. The dotted lines indicate the 148-Bq/m$$^3$$ US limit for indoor radon concentration [3]. The top figure (Atotsu tunnel) starts around 7 April 2016 as the detector was not deployed in the mine beforehand. The error bars include statistical uncertainties, as well as the uncertainty on the detector’s correction factor. Fig. 15. View largeDownload slide Correlation between the hourly averaged radon concentration measured in the Atotsu tunnel and the speed of the wind in this tunnel. Positive values indicate an air flow coming from outside of the mine and negative values indicate the opposite. The color scale shows the number of measurements in each bin. Fig. 15. View largeDownload slide Correlation between the hourly averaged radon concentration measured in the Atotsu tunnel and the speed of the wind in this tunnel. Positive values indicate an air flow coming from outside of the mine and negative values indicate the opposite. The color scale shows the number of measurements in each bin. In fall (October–November) and spring (April), daily fluctuations of the radon concentration are observed. Figure 16 illustrates this phenomenon. The radon concentration typically increases during daytime (usually from noon) and decreases during nighttime. Our interpretation of this phenomenon is that the temperature outside the mine fluctuates around the value leading to a change of the wind direction in the tunnel. Fig. 16. View largeDownload slide Hourly averaged radon concentration (bottom) and wind velocity (top) measured in the Atotsu tunnel (detector #1) from 8–16 April. Fig. 16. View largeDownload slide Hourly averaged radon concentration (bottom) and wind velocity (top) measured in the Atotsu tunnel (detector #1) from 8–16 April. 6.2. Radon concentration in the fresh air As shown in the previous section, the radon concentration in the air in the mine is very high. Therefore, to reduce the exposure of workers and experiments, fresh air needs to be brought into the experimental areas. This fresh air is brought through a 1.8-km pipe [4] going through the Atotsu tunnel. This pipe brings $$100$$ m$$^3$$ of fresh air per minute into the mine. The humidity in the pipe is kept between $$2.5$$ and $$7$$ g/m$$^{3}$$ by a dryer. We positioned two detectors to monitor the radon concentration in this pipe. One was placed at the input outside of the mine, in the so-called “Rn hut”, which houses a large compressor used to send the fresh air, and the other detector was placed at one of the outputs of the pipe in the Super-Kamiokande dome. Figure 17 shows the concentrations measured by these detectors from January 2016 to October 2017. A seasonal variation is observed in this figure, with an average radon concentration of $$50.7 \pm 1.0$$ Bq/m$$^{3}$$ during summer, and $$19.7 \pm 0.2$$ Bq/m$$^{3}$$ during the rest of the year. This observation suggests a feedback of the high-radon-concentration air from Atotsu tunnel. Figure 18 shows the hourly averaged radon concentration for both detectors over August 2016 and suggests that both detectors measured the same concentration and that there are no leaks in the pipe (i.e., no additional radon coming from the air in the mine). Fig. 17. View largeDownload slide 2-day-averaged radon concentration measured at the input (top) and output (bottom) of the fresh-air pipe. The detector at the input of the fresh-air pipe was deployed at the end of June 2016, hence the absence of data beforehand. The error bars include statistical uncertainties, as well as the uncertainty on the detector’s correction factor. Fig. 17. View largeDownload slide 2-day-averaged radon concentration measured at the input (top) and output (bottom) of the fresh-air pipe. The detector at the input of the fresh-air pipe was deployed at the end of June 2016, hence the absence of data beforehand. The error bars include statistical uncertainties, as well as the uncertainty on the detector’s correction factor. Fig. 18. View largeDownload slide Hourly averaged radon concentration measured at the input (plain circles, black) and output (red) of the fresh-air pipe. Fig. 18. View largeDownload slide Hourly averaged radon concentration measured at the input (plain circles, black) and output (red) of the fresh-air pipe. Figure 18 also indicates variations between daytime and nighttime radon concentrations in the fresh air. Such variations were observed over the whole year. During a typical day, the radon concentration tends to increase during the nighttime and decrease during the daytime. This phenomenon comes from the stratification of the atmosphere during the nighttime, and from the convection due to the Sun’s radiation during the daytime. 6.3. Radon concentration in the Super-Kamiokande area As shown in Fig. 12, several radon detectors are settled in the Super-Kamiokande area. Figure 19 shows the radon concentration measured in the dome of the Super-Kamiokande experiment from January 2016 to October 2017. The positions of the detectors in the Super-Kamiokande area are indicated in Fig. 13. The humidity in this area ranges from 6–8 g/m$$^{3}$$. The periods without data points are periods at which the detectors were not running due to power cuts or maintenance or because the detector was yet to be deployed. The concentration plots show the same patterns of radon-concentration variation, indicating that they are in the same atmosphere. The differences in the absolute value of the radon concentration can be explained by the positions of these detectors. In this dome, from the entrance of the Super-Kamiokande area to the opposite side, the concentrations are shown by the figures (a), (b), (c), and (d) (detectors #5, #6, #7, and #8, respectively). The radon concentration measured near the tunnel side of the dome is lower than the one measured on the other side of the dome. This phenomenon is due to the position of the radonless air output, which is located on this side of the dome. The radon concentration in the air inside the tank of Super-Kamiokande is not affected by the variation of radon concentration in the dome [23]. Fig. 19. View largeDownload slide 2-day-averaged radon concentration measured in the Super-Kamiokande dome. The dotted line indicates the 148-Bq/m$$^3$$ US limit for indoor radon concentration [3]. The error bars include statistical uncertainties, as well as the uncertainty on the detector’s correction factor. Detectors #5, #6, and #8 were deployed in April 2016, hence the absence of data before this month. Fig. 19. View largeDownload slide 2-day-averaged radon concentration measured in the Super-Kamiokande dome. The dotted line indicates the 148-Bq/m$$^3$$ US limit for indoor radon concentration [3]. The error bars include statistical uncertainties, as well as the uncertainty on the detector’s correction factor. Detectors #5, #6, and #8 were deployed in April 2016, hence the absence of data before this month. A detector is settled in the entrance room of the Super-Kamiokande dome (detector #9). This room is closed by two iron doors, and acts as a transfer airlock between the tunnel and the dome, in order to prevent the radon-rich air from the tunnel entering the dome. Figure 20(a) shows the 2-day-averaged concentration measured by this detector from January 2016 to October 2017. Figure 20(b) shows the hourly averaged concentration measured by the detector in May 2016. Fig. 20. View largeDownload slide Radon concentration measured in the entrance room of the Super-Kamiokande dome (detector #9). The dotted line indicates the 148-Bq/m$$^3$$ US limit for indoor radon concentration [3]. The error bars include statistical uncertainties, as well as the uncertainty on the detector’s correction factor. Fig. 20. View largeDownload slide Radon concentration measured in the entrance room of the Super-Kamiokande dome (detector #9). The dotted line indicates the 148-Bq/m$$^3$$ US limit for indoor radon concentration [3]. The error bars include statistical uncertainties, as well as the uncertainty on the detector’s correction factor. Daily fluctuation of the radon concentration is observed. The radon concentration increases during daytime, and decreases during nighttime. This phenomenon is due to the radon-rich air from the tunnel entering the room when workers (including shifters) enter the Super-Kamiokande area. During nighttime the radon concentration decrease as the injection of radon-rich air from the tunnel is reduced. Confirmation of this interpretation arises from the observation of the low radon concentration during Japanese national holidays (such as 3–5 May in Fig. 20(b)), and on Saturdays and Sundays, when few workers enter the area. On 20 October 2016, a sudden increase in the radon concentration can be observed in Fig. 20(a), as well as in Fig. 19. The door between Atotsu tunnel and the entrance room was left open overnight while the radon concentration in the tunnel was about $$1700$$ Bq/m$$^{3}$$, as shown in Fig. 14(a). In Figs. 19 and 20(a), another radon concentration increase is observed in February 2017, due to an interruption of the aeration system of the entrance room. 7. Summary New 1-L radon detectors and Raspberry Pi electronics have been developed for radon monitoring at the Kamioka Observatory. As of December 2017, twenty-one 1-L detectors using Raspberry Pi electronics are being used to monitor the radon concentration in the air in the mine. A total of 28 of these devices are used within the research facility (the Kamioka mine and surface buildings). These detectors and electronics have now been used for more than one year with stable data acquisition, as summarized in Table 2. We observed seasonal variations of the radon concentration in the air of the mine. Day–night variations of the radon concentration in the fresh air brought into the mine have also been observed. Table 2. Summary table of the average radon concentrations in different important areas of the mine. The second row indicates the number of detectors measuring the atmosphere of each area. The third row indicates the yearly averaged radon concentration. The fourth to seventh rows indicate the four seasonally averaged radon concentrations. Winter is defined as being between 21 December and 21 March; spring between 21 March and 21 June; summer between 21 June and 21 September; and fall between 21 September and 21 December.  Area Super-K dome Super-K entrance Atotsu tunnel Fresh-air pipe  Detectors 4 1 1 2 Averaged Year $$64.2 \pm 0.3$$ $$137.7 \pm 1.2$$ $$1082.6 \pm 10.5$$ $$28.6 \pm 0.3$$ radon Winter $$56.0 \pm 0.6$$ $$\,\,72.1 \pm 1.3$$ $$\,\,209.6 \pm 5.0$$ $$14.1 \pm 0.3$$ conc. Spring $$63.8 \pm 0.6$$ $$146.9 \pm 2.6$$ $$1261.1 \pm 23.3$$ $$21.5 \pm 0.5$$ (Bq/m$$^3$$) Summer $$79.2 \pm 0.8$$ $$194.4 \pm 3.5$$ $$1739.1 \pm 32.1$$ $$50.7 \pm 1.0$$ Fall $$55.2 \pm 0.5$$ $$136.0 \pm 2.5$$ $$\,\,791.0 \pm 14.3$$ $$22.7 \pm 0.4$$  Area Super-K dome Super-K entrance Atotsu tunnel Fresh-air pipe  Detectors 4 1 1 2 Averaged Year $$64.2 \pm 0.3$$ $$137.7 \pm 1.2$$ $$1082.6 \pm 10.5$$ $$28.6 \pm 0.3$$ radon Winter $$56.0 \pm 0.6$$ $$\,\,72.1 \pm 1.3$$ $$\,\,209.6 \pm 5.0$$ $$14.1 \pm 0.3$$ conc. Spring $$63.8 \pm 0.6$$ $$146.9 \pm 2.6$$ $$1261.1 \pm 23.3$$ $$21.5 \pm 0.5$$ (Bq/m$$^3$$) Summer $$79.2 \pm 0.8$$ $$194.4 \pm 3.5$$ $$1739.1 \pm 32.1$$ $$50.7 \pm 1.0$$ Fall $$55.2 \pm 0.5$$ $$136.0 \pm 2.5$$ $$\,\,791.0 \pm 14.3$$ $$22.7 \pm 0.4$$ View Large Table 2. Summary table of the average radon concentrations in different important areas of the mine. The second row indicates the number of detectors measuring the atmosphere of each area. The third row indicates the yearly averaged radon concentration. The fourth to seventh rows indicate the four seasonally averaged radon concentrations. Winter is defined as being between 21 December and 21 March; spring between 21 March and 21 June; summer between 21 June and 21 September; and fall between 21 September and 21 December.  Area Super-K dome Super-K entrance Atotsu tunnel Fresh-air pipe  Detectors 4 1 1 2 Averaged Year $$64.2 \pm 0.3$$ $$137.7 \pm 1.2$$ $$1082.6 \pm 10.5$$ $$28.6 \pm 0.3$$ radon Winter $$56.0 \pm 0.6$$ $$\,\,72.1 \pm 1.3$$ $$\,\,209.6 \pm 5.0$$ $$14.1 \pm 0.3$$ conc. Spring $$63.8 \pm 0.6$$ $$146.9 \pm 2.6$$ $$1261.1 \pm 23.3$$ $$21.5 \pm 0.5$$ (Bq/m$$^3$$) Summer $$79.2 \pm 0.8$$ $$194.4 \pm 3.5$$ $$1739.1 \pm 32.1$$ $$50.7 \pm 1.0$$ Fall $$55.2 \pm 0.5$$ $$136.0 \pm 2.5$$ $$\,\,791.0 \pm 14.3$$ $$22.7 \pm 0.4$$  Area Super-K dome Super-K entrance Atotsu tunnel Fresh-air pipe  Detectors 4 1 1 2 Averaged Year $$64.2 \pm 0.3$$ $$137.7 \pm 1.2$$ $$1082.6 \pm 10.5$$ $$28.6 \pm 0.3$$ radon Winter $$56.0 \pm 0.6$$ $$\,\,72.1 \pm 1.3$$ $$\,\,209.6 \pm 5.0$$ $$14.1 \pm 0.3$$ conc. Spring $$63.8 \pm 0.6$$ $$146.9 \pm 2.6$$ $$1261.1 \pm 23.3$$ $$21.5 \pm 0.5$$ (Bq/m$$^3$$) Summer $$79.2 \pm 0.8$$ $$194.4 \pm 3.5$$ $$1739.1 \pm 32.1$$ $$50.7 \pm 1.0$$ Fall $$55.2 \pm 0.5$$ $$136.0 \pm 2.5$$ $$\,\,791.0 \pm 14.3$$ $$22.7 \pm 0.4$$ View Large Footnotes 1http://www.win-ei.com. Acknowledgements The authors would like to thank Mitsuhiro Nakamura, from the Fundamental Particle Physics Laboratory, Graduate School of Science of Nagoya University, for his help in the development of the preamplifier circuit. We would like to express our sincere gratitude to Prof. Takao Iida, from the Graduate School of Engineering of Nagoya University, who gave us adequate advice and discussion on the development of the radon detector. We would also like to thank the Super-Kamiokande Collaboration for their help in conducting this study. We gratefully acknowledge the cooperation of the Kamioka Mining and Smelting Company. References [1] Raspberry Pi documentation . (Available at: https://www.raspberrypi.org/documentation/) [2] Brookhaven National Laboratory , National Nuclear Data Center . (Available at: http://www.nndc.bnl.gov/.) [3] EPA website . (Available at: https://www.epa.gov/radon/health-risk-radon/.) [4] Fukuda S. et al. [ Super-Kamiokande Collaboration ], Nucl. Instrum. Meth. Phys. Res. A 501 , 418 ( 2003 ). ( https://doi.org/10.1016/S0168-9002(03)00425-X ) Crossref Search ADS [5] Gando A. et al. [ KamLAND-Zen Collaboration ], Phys. Rev. Lett. 110 , 062502 ( 2013 ). ( https://doi.org/10.1103/PhysRevLett.110.062502 ) Crossref Search ADS PubMed [6] Abe K. et al. [ XMASS Collaboration ], Nucl. Instrum. Meth. Phys. Res. A ( https://doi.org/10.1016/j.nima.2013.03.059 ) 716 , 78 ( 2013 ). Crossref Search ADS [7] Umehara S. et al. , Phys. Rev. C 78 , 058501 ( 2008 ). ( https://doi.org/10.1103/PhysRevC.78.058501 ) Crossref Search ADS [8] Nakamura K. et al. , Prog. Theor. Exp. Phys. 2015 , 043F01 ( 2015 ). ( https://doi.org/10.1093/ptep/ptv041 ) Crossref Search ADS [9] Kotrappa P. , Dua S. K. , Gupta P. C. , and Mayya Y. S. , Health Phys. 41 , 35 ( 1981 ). ( http://dx.doi.org/10.1097/00004032-198107000-00004 ) Crossref Search ADS PubMed [10] Tasaka S. and Sasaki Y. , Radioisotopes 45 , 741 ( 1996 ). ( http://dx.doi.org/10.3769/radioisotopes.45.12_741 ) Crossref Search ADS [11] Iida T. , Y. Ikebe , T. Hattori , H. Yamanishi , S. Abe , K. Ochifuji , and S. Yokoyama , Health Phys . 54 , 139 ( 1988 ). ( https://www.ncbi.nlm.nih.gov/pubmed/3338911 ) Crossref Search ADS PubMed [12] Maxim MAX153 datasheet . (Available at: https://datasheets.maximintegrated.com/en/ds/MAX153.pdf.) [13] Texas Instruments ADC104S021 datasheet . (Available at: http://www.ti.com/lit/ds/symlink/adc104s021.pdf.) [14] Wiring Pi, GPIO Interface library for the Raspberry Pi . (Available at: http://wiringpi.com/.) [15] Sato T. et al. , Earth Planets Space 56 , BF03353071 ( 2014 ). ( http://doi.org/10.1186/BF03353071 ) [16] Chu K. D. and Hopke P. K. , Environ. Sci. Tech. 22 , 711 ( 1988 ). ( http://dx.doi.org/10.1021/es00171a016 ) Crossref Search ADS [17] VAISALA humidity conversion formulas , B210973EN-F ( 2013 ) (Available at https://www.vaisala.com/sites/default/files/documents/Humidity_Conversion_Formulas_B210973EN-F.pdf). [18] PYLON® manual A900071 . [19] Hosokawa K. , A. Murata , Y. Nakano , Y. Onishi , H. Sekiya , Y. Takeuchi , and S. Tasaka , Prog. Theor. Exp. Phys. 2015 , 033H01 ( 2015 ). ( http://dx.doi.org/10.1093/ptep/ptv018 ) Crossref Search ADS [20] Currie L. A. , Anal. Chem. 40 , 586 ( 1968 ). ( http://dx.doi.org/10.1021/ac60259a007 ) Crossref Search ADS [21] Brun R. and Rademakers F. , Nucl. Instrum. Meth. Phys. Res. A 389 , 81 ( 1997 ). ( http://dx.doi.org/10.1016/S0168-9002(97)00048-X ) Crossref Search ADS [22] Takeuchi Y. , K. Okumura , T. Kajita , S. Tasaka , H. Hori , M. Nemoto , and H. Okazawa , Nucl. Instrum. Meth. Phys. Res. A 421 , 334 ( 1999 ). ( http://dx.doi.org/10.1016/S0168-9002(98)01204-2 ) Crossref Search ADS [23] Nakano Y. et al. , Nucl. Instrum. Meth. Phys. Res. A 867 , 108 ( 2017 ). ( https://doi.org/10.1016/j.nima.2017.04.037 ) Crossref Search ADS © The Author(s) 2018. Published by Oxford University Press on behalf of the Physical Society of Japan. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited. TI - Development of new radon monitoring systems in the Kamioka mine JF - Progress of Theoretical and Experimental Physics DO - 10.1093/ptep/pty091 DA - 2018-09-01 UR - https://www.deepdyve.com/lp/oxford-university-press/development-of-new-radon-monitoring-systems-in-the-kamioka-mine-pmbALpjghC VL - 2018 IS - 9 DP - DeepDyve ER -