TY - JOUR AU - Miedema,, Kor AB - Abstract A reference method that specifically measures hemoglobin (Hb) A1c is an essential part of the reference system for the international standardization of Hb A1c/glycohemoglobin. We have developed a new method for quantification, based on the specific N-terminal residue of the hemoglobin β-chains. Enzymatic cleavage of the intact hemoglobin molecule with endoproteinase Glu-C has been optimized to obtain the β-N-terminal hexapeptides of Hb A1c and Hb A0. These peptides have been separated by reversed-phase HPLC and quantitated by electrospray ionization-mass spectrometry (method A) or by capillary electrophoresis (method B). With these peptides and hyphenated separation techniques, it has been possible to overcome the insufficient resolution of currently used protein separation systems for Hb A1c. Hemoglobin (Hb)1 A1c is the major glycohemoglobin species in human blood (1)(2) and has been used for ∼20 years for long-term assessment of glycemic control in diabetic patients. The comprehensive Diabetes Control and Complications Trial (DCCT) has provided ample evidence that microvascular complications such as retinopathy, nephropathy, and neuropathy are directly related to the degree of hyperglycemia in patients with insulin-dependent diabetes and that measurement of Hb A1c in blood is an excellent tool for long-term monitoring of the glycemic state of diabetic patients (3)(4)(5)(6). The DCCT study has also clearly demonstrated the need for reliable and reproducible measurements of Hb A1c because only a relatively small difference exists between the average Hb A1c concentrations in patients at low and high risk of developing the dreaded late complications. Many different routine methods (>20) claiming to measure Hb A1c are currently used by the clinical laboratories. The methods either exploit the charge differences existing between Hb A1c and Hb A0 (ion-exchange chromatography, electrophoresis, isoelectric focusing) or are immunological methods that use poly- or monoclonal antibodies directed towards the glycated N-terminal group of the β-chain of Hb. At present, no internationally accepted reference system, to which the routine assays could be adjusted (7), exists nor do any internationally approved primary or secondary reference materials or a reference method. As a result, values differ in a clinically unacceptable magnitude between methods and laboratories as has been demonstrated in several trials (8)(9)(10)(11)(12). To overcome this problem, national standardization activities have been initiated in some countries to achieve standardization on a national scale, but these activities are independent of each other and tend to use different approaches. To achieve a uniform international standardization, the IFCC has established a working group on Hb A1c standardization that coordinates activities worldwide. The working group is developing a reference system that will be the basis for the international standardization (13). The first step was a clear definition of the analyte Hb A1c on the basis of its molecular structure. Historically Hb A1c was defined as a certain peak in an HPLC system, but this is no longer scientifically acceptable. To overcome this problem, one suggested definition for Hb A1c was Hb that is glycated at the N-terminal of the β-chain [β-(N-deoxyfructosyl)hemoglobin] because this is the major compound in the HPLC peaks and the major form of all glycohemoglobins; however, the issue of potential double glycation at both β-chains or an additional glycation at any lysine residue was not discussed explicitly (2). To make this clear, the IFCC group has now defined Hb A1c as Hb that is irreversibly glycated at one or both N-terminal valines of the β-chains. This also covers Hb that is additionally glycated at any lysine residue in the β-chain. Hb that is solely glycated at a lysine residue is not considered to be Hb A1c. The second step of the standardization process is the development of a reference method that can measure Hb A1c specifically according to this molecular definition. At this time, such a method is not yet available. HPLC methods, which have often been used as reference methods for standardization of routine tests, are not really reference methods in the scientific sense. They provide precision and long-term stability but lack specificity. Different values for Hb A1c in the same blood are obtained depending on the conditions used, because the Hb A1c peaks contain different kinds and amounts of substances that are not Hb A1c[14–19]. The differences between certain methods can be rather large. Between the Bio-Rex 70 HPLC method used as a designated comparison method for the internal standardization of the DCCT study (2) and the Mono S ion-exchange chromatography method (20)(21), differences as large as 20% were found at the reference range concentrations, despite both methods’ claiming to measure Hb A1c [22]. Although the Mono S system is more specific than the Bio-Rex 70 method, it cannot be accepted as a reference method because carbamylated Hb and dimers of glycated α- and nonglycated β-chains coelute with Hb A1c in this system (19)(23). Because none of the currently available methods for Hb A1c determination shows sufficient specificity to be used as a reference method, we have developed one that distinguishes the smaller N-terminal parts of the β-chains of Hb A1c from those of the Hb A0 molecules, thus avoiding the heterogeneity created by modifications of other glycation sites of the Hb molecule. The most specific Reference Methods in clinical chemistry available today are based on mass spectrometric detection (gas chromatography–mass spectrometry with isotopic dilution). Most were developed in the 1980s, e.g., for cholesterol, progesterone, testosterone, cortisol, estradiol, or thyroxine (24). We have applied electrospray ionization (ESI-MS), which is a relatively new mass spectrometric technique suited for the analysis of polar biomolecules of low and high molecular mass (25). The process of ESI is characterized by formation of polyprotonated or polydeprotonated ions of intact analyte molecules that are produced from a fine spray of an aqueous solution of the analyte, assisted by a strong electrical field at atmospheric pressure. The ions have low internal energy and are not prone to fragmentation. The measurement of mass-to-charge values of these multiple charged ions gives the inherent molecular mass of the analyte molecules. The ion currents produced by ESI are dependent on the analyte concentration. The ESI technique has improved dramatically within the last 5 years, and detection limits in the lower picomole to upper femtomole range are now available. ESI-MS is ideally suited for an on-line coupling with HPLC, and the combination of these techniques might become a powerful instrument for the development of new Reference Methods for a variety of analytes. Considering the high cost of ESI-MS equipment, we concurrently investigated the possibility of developing and providing another method based on capillary electrophoresis (CE), a technique that is more generally available in clinical laboratories. Materials and Methods preparation of blood samples Fresh blood samples collected in EDTA-containing tubes (Becton Dickinson) were used. Mixed blood (1.5 mL) was washed twice with 10 mL of saline and centrifuged for 10 min at 3000g. The sedimented cells were incubated in 10 mL of saline at 37 °C for 4 h to eliminate pre-Hb A1c. Hemolysate was prepared by mixing the cells with 1.5 mL of water, followed by centrifugation for 20 min at 3000g. Supernatants were mixed with equal parts of buffer, 50 mmol/L 2-(N-morpholino)propanesulfonic acid), 1 mmol/L EDTA, 10 mmol/L potassium cyanide, and 0.2 g/L sodium azide. The pH was adjusted to 6.2 with 2 mol/L hydrochloric acid. Cell debris was removed by centrifugation before storage of a clear hemolysate at −70 °C. enzymatic cleavage of hemoglobin Endoproteinase Glu-C sequencing grade (Boehringer Mannheim; cat. no. 1047817) was used for cleavage of Hb. An aliquot of 1 mg of Hb from the patients’ samples or the primary calibrators was transferred to a 1-mL crimp cap glass vial, mixed with 20 μL of a solution of 50 μg of Glu-C in 250 μL of water and diluted with 25 mmol/L ammonium acetate buffer (pH 4.0) up to a total volume of 500 μL. Vials were carefully closed with crimp caps and incubated with gently shaking at 37 °C for 18 h. Digestion was stopped by freezing at −20 °C. For the cleavage experiments with trypsin an Hb solution consisting of 2 mg of Hb in 300 μL of distilled water was used. After 3 min of boiling, 200 μL of ammonium bicarbonate (0.5 mol/L, pH 8.2) was added. The solution was incubated for 2 h at 37 °C with 20 μL of N-tosyl-l-phenylalanyl chloromethyl ketone-treated trypsin (Sigma, 5 mg/mL in 1 mmol/L HCl and 12 mmol/L CaCl2). The digestion was stopped by adding 2 drops of 6 mol/L HCl. calibrators Human blood from healthy, nondiabetic volunteers was used to isolate Hb A0 and Hb A1c for the preparation of primary calibrators. Erythrocytes were sedimented by centrifugation, washed with sodium chloride, hemolyzed with water, and stabilized in an EDTA/KCN solution to isolate hemoglobins. Purification was achieved by three chromatographic steps: cation-exchange chromatography on SP-Sepharose HP, affinity chromatography on GLYCO-GEL® II Boronate Affinity Gel, and cation-exchange chromatography on SP-Sepharose HP again. Purified Hb A0 and Hb A1c were stored stabilized by KCN in an EDTA/2-(N-morpholino)propanesulfonic acid buffer system. Purified Hb A1c and Hb A0 were analytically characterized by Mono S cation-exchange chromatography, capillary isoelectric focusing, ESI-MS, and quantification of β-N-terminal peptides by enzymatic cleavage and reversed-phase HPLC. Total Hb was determined by the Hb cyanide method (26). Details of the preparation and characterization are described elsewhere (23). Calibrators were prepared by mixing Hb A1c and Hb A0 primary materials. Mixing of calibrators was done with calibrated pipettes, on the basis of weight according to the total Hb concentration and the degree of Hb A0 impurity in the Hb A1c preparation. For calibration of the analytical systems described below, a set of 6 calibrators covering the range of 0–15% Hb A1c (relative to total Hb) was used. method a. hplc separation and on-line esi-ms detection Samples prepared as described above were analyzed in a combined HPLC-ESI-MS system. The HPLC system consisted of an HP 1090 liquid chromatograph (Hewlett-Packard) with a DR 5 solvent-delivery system, a thermostat-equipped autosampler, an autoinjector, a Rheodyne no. 7010P (ERC) column-switching valve, a 0.159-cm Swagelok no. SS-100-3 (B.E.S.T.) T-piece, a relay box for control of pneumatic valves (Festo), a Kratos Programmable Absorbance Detector Spectroflow 783 (Bioanalytische Instrumente) with 2.4-μL cell volume, and a 2.1 × 150 mm analytical HPLC column (ZORBAX SB-CN, 5 μm, no. AS-RT-1245; P/N:883700.905; Axel Semrau). Flow rate was set to 300 μL/min, column temperature to 50 °C. Injection volume was 50 μL of digest. A gradient elution was performed with eluent A (0.25 mL/L trifluoroacetic acid in water) and eluent B (0.23 mL/L trifluoroacetic acid in acetonitrile): 0 min, 0% B; 30 min, 15% B; 31 min, 100% B; 34 min, 100% B; 35 min, 0% B; 39 min, stop run). For quantitative determinations a column switching valve was positioned after the HPLC column, and only the fraction between the 5-min and 20-min elution times was allowed to enter the detection system to avoid contamination of the electrospray ion source. The photometric detector was set to 214 nm and was used for control purposes only. The mass spectrometric system was an SSQ700 single-stage quadrupole mass spectrometer with an electrospray ion source (Finnigan MAT). The HPLC system was connected on-line with the photometric detector and the electrospray ion source by 0.12-mm (i.d.) steel capillaries. The electrospray ion source was run with 414 kPa nitrogen sheath gas and nitrogen auxiliary gas at an HPLC flow rate of 300 μL/min. Spray voltage was 4.5 kV, transfer capillary temperature 200 °C. The mass spectrometer was tuned and calibrated with tetrapeptide MRFA, myoglobin mixture; resolution was set to 0.7 amu peak half-width and the electron multiplier to 13 kV. Acquisition mode was set to centroid, multiple ion detection at m/z 348.2 and 429.2 for the double-protonated ions of nonglycated and glycated N-terminal hexapeptides of the hemoglobin β-chain. For measurements a sequence was set up for calibration bracketing: primary calibrators–samples–primary calibrators. Four injections were made for each vial. The ion chromatograms for m/z 348.2 and 429.2 were recorded, the peak areas integrated, and the ratios of area m/z 429.2 vs m/z 348.2 calculated. The mean values of four injections at a time per vial were calculated. The ratio of peak areas was a linear function of the percent Hb A1c for a set of primary calibrators. With the data set of primary calibrators a calibration function was calculated by linear regression of the measured ratio against the default values of percent Hb A1c. This calibration function was used to determine the percent Hb A1c values of the measured patient samples. method b: hplc and ce Samples prepared as described above were separated by reversed-phase chromatography on a C18 column (PepRPC HR 5/5; Pharmacia). HPLC separation was done as described earlier (27) with the following modifications: an HPLC system, including two P-500 pumps, a GP-250 gradient programmer, and a UV detector 2141 (LKB-Pharmacia); flow rate was 1 mL/min, gradient elution was performed with eluent A (1 mL/L trifluoroacetic acid in water) and eluent B (1 mL/L trifluoroacetic acid in acetonitrile): 0 min, 0% B; 40 min, 18% B; 41 min, 100% B; 43 min, 100% B; 44 min, 0% B; 52 min, stop run. Absorbance was monitored at 214 and 280 nm. The actual fraction, the first dominating peak, was identified by amino acid analysis, collected, and lyophilized. The material was dissolved in 40 μL of 0.1 mL/L trifluoroacetic acid in water immediately before electrophoresis. In a second step the mixture of glycated and nonglycated β-N-terminal hexapeptides was separated by CE performed on a Beckman P/ACE system 5000 equipped with System GoldTM version 8.1 software. Fused-silica capillaries, 50 μm (i.d) × 67 cm (no. 338472; Beckman Instruments), were used. New capillaries were conditioned by rinsing with 0.1 mol/L NaOH for 10 min and buffer for another 10 min. A typical run would be 100 mmol/L phosphate buffer, pH 2.5, 15-s pressure injection, 25-kV voltage, temperature at 20 °C, and absorbance monitored at 214 nm. Calibration was carried out as described in method A. hplc comparison method Mono S ion-exchange chromatography, a reliable and well-established routine method, was chosen for a comparison with the newly developed methods. In this method Hb A1c is separated from other hemoglobins Hb A1a, Hb A1b, Hb F, acetylated Hb, α-chain dimers, the glutathione adduct Hb A3, and Hb A0 by a salt gradient in malonate buffer at pH 5.7 on a Mono STM HR 5/5 column (Pharmacia) as described earlier (20)(21). The pre-Hb A1c is eliminated by incubation of samples in a citrate/phosphate buffer, pH 5.4, for 30 min at 37 °C (21). method comparison studies Two sets of patient samples were measured on different occasions with the HPLC-MS procedure, the HPLC-CE procedure, and Mono S ion-exchange chromatography. The results were statistically evaluated. Results enzymatic cleavage of hemoglobin The enzyme endoproteinase Glu-C cleaves the hemoglobin molecule so that the N-terminal hexapeptides C4H9O4-CO-CH2-NH-Val-His-Leu-Thr-Pro-Glu-COOH from Hb A1c and NH2-Val-His-Leu-Thr-Pro-Glu-COOH from Hb A0 were released from the β-chains. The digestion conditions were optimized to get high reproducibility and equal digestion kinetics for both Hb A0 and Hb A1c β-chains. A patient’s sample was digested over 5–40 h to study the digestion kinetics, and aliquots were taken every 4 h. The reaction was stopped by freezing. Under the chosen digestion conditions the ratio of released glycated and nonglycated hexapeptides was constant over a period of 40 h (Fig. 1 ). For the final methods we used 18-h digestion at 37 °C. method a: hplc separation and on-line esi-ms detection The HPLC separation has been optimized for the HPLC-ESI-MS system to achieve good resolution between glycated and nonglycated β-N-terminal hexapeptides and good separation from all other peptide fragments. Mass spectra from synthetic glycated and nonglycated β-N-terminal hexapeptide calibrators are shown in Fig. 2 . The spectra show both single- and double-protonated ions, and no fragmentation is observed. For quantitative measurements the double-protonated ions were chosen because of their better response behavior. Resolution of the mass analyzer was set to 1 Da. The very high specificity of the mass spectrometric detection is shown in Fig. 3 . The chromatogram recorded in a scan mode, which is similar to a photometric detection at 215 nm, is compared with the multiple-ion detection mode for the doubly protonated ions at m/z 348.2 and 429.2, which represent the hexapeptides released from Hb A1c and Hb A0. This comparison clearly shows the superior specificity of the MS detection. The detection limit of the analytical system at multiple-ion detection mode is sufficient to get superior signal-to-noise ratios. By simple least-squares regression a linear calibration function was generated (Fig. 4 ). The range over which the response (ratio of glycated hexapeptide to nonglycated hexapeptide) is linearly proportional to the percent Hb A1c was evaluated by measuring a set of six primary calibrators covering the range 0–15% Hb A1c. The stability and reproducibility of the total analytical system were proven by repeated measurement of three patients’ samples (∼4%, 5%, and 7.5% Hb A1c) within a period of 4 weeks (duplicates on 4 different days). The three samples were run together with a larger set of patients’ samples. Including all calibration and test assays, >250 injections were done during this period, in addition to the routine use of the mass spectrometer for qualitative peptide and protein analysis. The overall CVs for Hb A1c amounts of 4.09%, 5.37%, and 7.91% were 2.0%, 1.2%, and 2.5%, respectively. This demonstrates a very good reproducibility of the system. The overall procedure of the complete method is shown in the flow chart in Fig. 5 . method b: hplc separation and ce We also studied a second independent multidimensional approach to quantify the ratio of glycated to nonglycated β-N-terminal hexapeptides. This was done by reversed-phase HPLC combined with off-line CE and photometric detection. Glycated and nonglycated β-N-terminal hexapeptides coeluted together on the C18 column used. This step was used for an enrichment of these peptides. In a second step, CE, which separated the C18 fraction into two peaks, was introduced (Fig. 6 ). By comparing synthetic glycated and nonglycated β-N-terminal hexapeptides, it was obvious that the peptides showed different absorbance values at 214 nm. The system was therefore calibrated in the same way as the ESI-MS method with the same calibrators (area % used for calculations). The within-day CVs of the HPLC-CE system for Hb A1c at 3.86%, 6.36%, and 12.0% were 1.92%, 1.58%, and 1.62%, respectively. The overall procedure of the complete method is shown in the flow chart in Fig. 5 . method comparison studies These studies were carried out to compare the two newly developed methods and also to compare the new methods with a reliable well-established routine HPLC method, the Mono S method. The comparison between the HPLC-CE and HPLC-MS methods (Fig. 7 ) shows an excellent agreement. The comparisons of the Mono S method against HPLC-MS also demonstrate good correlations (Fig. 8 ). Comparison of Mono S vs HPLC-CE is similar (y = 0.836 x + 1.35, r = 0.994). The remarkable slope is expected because of the theoretical considerations that Mono S gives higher Hb A1c values in the lower concentration range, because of impurities that coelute with the Hb A1c peak, and gives lower values in the upper concentration range, because of the additional glycation of ε-amino groups of lysine that will not coelute with the original Hb A1c peak, thus resulting in a reduction of the Hb A1c signal. Because all ion-exchange chromatographic systems are traceable to each other (9)(10), there will be traceability between currently used ion-exchange systems and the described candidate Reference Methods. Discussion Because the currently available methods for the measurement of Hb A1c in human blood do not fulfill the high requirements necessary for a Reference Method, we searched for a new principle that could result in a more specific method. Many analytical problems lie in the complexity of the Hb molecule and the multitude of glycohemoglobin species and other Hb adducts that exist in human blood. Because Hb A1c is defined as an Hb that is glycated at the N-terminal of the β-chain (3), we carried out studies to see whether it is possible to change from the determination of the total molecule to the measurement of the glycated and nonglycated β-N-terminal peptides of Hb. Therefore we studied the possibilities for an appropriate enzymatic cleavage of the Hb molecule. Trypsin, which cleaves peptide bonds with lysine and arginine at the C-terminal side, is a commonly used enzyme for peptide mapping in hemoglobinopathies. We were able to show that the lysine residues at position 8 in the β-chain are glycated in samples with increased Hb A1c concentrations but not in those with normal concentrations. Thus the use of trypsin to release N-terminal octapeptides for quantification purposes would include the risk of getting doubly or singly glycated octapeptides at the Lys-8 position. Therefore trypsin cleavage was not usable. Endoproteinase Glu-C cleaves the N-terminal part of the β-chain between the two glutamic acid residues at positions 6 and 7. The resulting fragments contain only a single glycation site at the N-terminal valine and can thus be used to separate Hb A1c. The actual cleaving site is easily exposed to the enzyme under mild denaturing conditions at pH 4.0. Complete denaturation before digestion exposes additional substrates to the enzyme and yields a more complex peptide mixture. With the modern multidimensional analytical techniques of on-line HPLC and ESI-MS or the off-line system of HPLC and CE, the two β-N-terminal hexapeptides of Hb A1c and Hb A0 could be separated and quantified with the necessary analytical performance. By analyzing the mixture of peptide fragments resulting from the endoproteinase Glu-C digestion of whole-blood samples, we obtained high specificity and a low detection limit. The approach excludes the interference of carbamylated and acetylated N-terminal species as well as other N-terminal adducts or nonglycated β-chains from the dimer of glycated α-chain–nonglycated β-chain (which coelute for example with Hb A1c in the Mono S system). We have developed two quantitative analytical methods that specifically measure the ratio of the N-terminal hexapeptides of Hb A1c and Hb A0. The calibration of both methods was done with primary calibrators that are mixtures of Hb A1c and Hb A0 primary calibrators. The CE method is a simple and robust technique but requires an off-line preparation of samples. The ESI-MS method requires advanced equipment but permits highly specific detection like the classical Reference Methods in clinical chemistry and has the practical advantage of on-line coupling without time-consuming sample processing. In our method, MS has been used for the first time in a Reference Method for proteins or peptides. The method is a good example of how to adapt the well-known concept of Reference Methods for low-Mr analytes to high-Mr proteins. From a theoretical point of view HPLC-ESI-MS offers higher specificity than HPLC-CE. This motivates its position as the most advanced candidate for the forthcoming Hb A1c Reference Method. However, with the observed analytical performance, the HPLC-CE method could also serve as a Reference Method. According to the NCCLS definition of analytical methods, the HPLC-CE system could be classified as a Reference Method while the HPLC-ESI-MS method has the potential to be a Definitive Method. Both methods will now be evaluated further in an international network of reference laboratories organized by the IFCC Working Group on Hb A1c Standardization. Clearly, however, the Hb A1c values found with the new Reference Methods will be considerably lower than those found with many current routine methods, especially when they are calibrated against the DCCT method (3). The designated comparison method in that study was a cation-exchange HPLC method with Bio-Rex 70 as resin. Although the method was very stable during the 9-year period of the DCCT study, it has become evident that the Hb A1c result by this method is only 60% specific, due to β-N-deoxyfructosyl Hb (17). More sophisticated assays with other resins, e.g., Mono S, yield lower Hb A1c results, but still show a perfect correlation with the Bio-Rex 70 method (19)(20). The new Reference Methods will solve this lack of specificity. For this reason lower values of Hb A1c will be expected in nondiabetic persons as well as in diabetic patients. Therefore, reference ranges from nondiabetics and recommended values for optimal therapy in diabetic patients will have to be revised and adjusted to the new accuracy. Preliminary experiments with more specific methods, e.g., the Mono S method (18), have already shown that translation of the Hb A1c values found in the DCCT study into values based on these more specific methods is feasible. The translation of Hb A1c values found in the DCCT and other important clinical trials into values that are based on the new reference system will be a task of the IFCC Working Group on Hb A1c Standardization (13) because the previous clinical experience must be maintained. 1 Nonstandard abbreviations: Hb, hemoglobin; DCCT, Diabetes Control and Complications Trial; ESI-MS, electrospray ionization-mass spectrometry; CE, capillary electrophoresis. Figure 1. Open in new tabDownload slide Fig. 1. Digestion of hemolysates by endoproteinase Glu-C: formation of glycated and nonglycated β-N-terminal hexapeptides of Hb and their ratios. Figure 1. Open in new tabDownload slide Fig. 1. Digestion of hemolysates by endoproteinase Glu-C: formation of glycated and nonglycated β-N-terminal hexapeptides of Hb and their ratios. Figure 2. Open in new tabDownload slide Fig. 2. Electrospray mass spectra of glycated and nonglycated β-N-terminal hexapeptides. +H+, singly protonated peptide; +2H+, doubly protonated peptide. Figure 2. Open in new tabDownload slide Fig. 2. Electrospray mass spectra of glycated and nonglycated β-N-terminal hexapeptides. +H+, singly protonated peptide; +2H+, doubly protonated peptide. Figure 3. Open in new tabDownload slide Fig. 3. Scan mode HPLC-ESI-MS and multiple ion detection chromatograms. Scan mode and selected ion monitoring on doubly protonated ions: m/z 348, β-N-terminal hexapeptide; m/z 429, Glc-β-N-terminal hexapeptide. Figure 3. Open in new tabDownload slide Fig. 3. Scan mode HPLC-ESI-MS and multiple ion detection chromatograms. Scan mode and selected ion monitoring on doubly protonated ions: m/z 348, β-N-terminal hexapeptide; m/z 429, Glc-β-N-terminal hexapeptide. Figure 4. Open in new tabDownload slide Fig. 4. Calibration curve for HPLC-ESI-MS with primary calibrators. Figure 4. Open in new tabDownload slide Fig. 4. Calibration curve for HPLC-ESI-MS with primary calibrators. Figure 5. Open in new tabDownload slide Fig. 5. Flow chart of methods A and B. Figure 5. Open in new tabDownload slide Fig. 5. Flow chart of methods A and B. Figure 6. Open in new tabDownload slide Fig. 6. HPLC and off-line capillary electrophoretic separation. A, HPLC chromatogram for separation on C18 column; B, chromatogram for CE separation. Figure 6. Open in new tabDownload slide Fig. 6. HPLC and off-line capillary electrophoretic separation. A, HPLC chromatogram for separation on C18 column; B, chromatogram for CE separation. Figure 7. Open in new tabDownload slide Fig. 7. HPLC-CE vs HPLC-ESI-MS. Figure 7. Open in new tabDownload slide Fig. 7. HPLC-CE vs HPLC-ESI-MS. Figure 8. Open in new tabDownload slide Fig. 8. Mono S vs HPLC-ESI-MS. Figure 8. Open in new tabDownload slide Fig. 8. Mono S vs HPLC-ESI-MS. The scientific part of this work was supported by the IFCC Working Group on Standardization of Hb A1c: K. Miedema, The Netherlands (Chairman); D. Goldstein, US; I. Goodall, Australia; W. Hoelzel, Germany; J.-O. Jeppsson, Sweden; R. Little, US; Th. J. Penders, The Netherlands; K. Shima, Japan; C. Weykamp, The Netherlands. We thank all members for valuable discussions and encouragement. We gratefully acknowledge the skillful technical assistance of Anna Arnetorp and Ann-Christine Martensson, the loan of capillary electrophoresis equipment by Beckman Instruments, and the financial support of Boehringer Mannheim. 1 Shapiro R, McManus MJ, Zolut C, Bunn HF. Sites of nonenzymatic glycosylation of human hemoglobin A. J Biol Chem 1980 ; 255 : 3120 -3127. PubMed 2 Goldstein DE, Little RR, Wiedmeyer H-M, England JD, McKenzie EM. Glycated hemoglobin: methodologies and clinical applications. Clin Chem 1986 ; 32 (Suppl): B64 -B70. PubMed 3 . The Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993 ; 329 : 977 -986. Crossref Search ADS PubMed 4 Santiago JV. Lessons from the Diabetes Control and Complications Trial. Diabetes 1993 ; 42 : 1549 -1554. Crossref Search ADS PubMed 5 Benjamin J, Sacks DB. Glycated protein update. Implications of recent studies, including the Diabetes Control and Complications Trial. Clin Chem 1994 ; 40 : 683 -687. Crossref Search ADS PubMed 6 Goldstein D, Little R, Wiedmeyer H-M, England JD, Rohlfing CL, Wilke AL. Is glycohemoglobin testing useful in diabetes mellitus? Lessons from the Diabetes Control and Complications Trial. Clin Chem 1994 ; 40 : 1937 -1940. 7 Bruns DE. Standardization, calibration, the care of diabetic patients [Editorial]. Clin Chem 1992 ; 38 : 2363 -2364. Crossref Search ADS PubMed 8 Little RR, Wiedmeyer HM, England JD, Wilke AL, Rohlfing CL, Wians FH, Jr, et al. Interlaboratory standardization of measurements of glycohemoglobins. Clin Chem 1992 ; 38 : 2472 -2478. Crossref Search ADS PubMed 9 Weykamp CW, Penders TJ, Muskiet FA, van der Slik W. Effect of calibration on dispersion of glycohemoglobin values determined by 111 laboratories using 21 methods. Clin Chem 1994 ; 40 : 138 -144. Crossref Search ADS PubMed 10 Weykamp CW, Penders TJ, Miedema K, Muskiet FA, van der Slik W. Standardization of glycohemoglobin results and reference values in whole blood studied in 103 laboratories using 20 methods. Clin Chem 1995 ; 41 : 82 -86. Crossref Search ADS PubMed 11 Shima K, Endo J, Oimomi M, Oshima I, Omori Y, Katayama Y, et al. Interlaboratory difference in HbA1c measurement in Japan. An interim report of the Committee on an Interlaboratory Standardization of HbA1c Determination, the Japan Diabetes Society. J Jpn Diabetes Soc 1994 ; 37 : 233 -243. 12 Mosca A, Carpinelli A, Bonini PA. Ethylene glycol-stabilized haemolysates for glycated haemoglobin assay. Further analytical investigations and applications to quality control programmes. J Clin Chem Clin Biochem 1987 ; 25 : 447 -450. PubMed 13 Hoelzel W, Miedema K. Development of a reference system for the international standardization of HbA1c/glycohemoglobin determinations. J IFCC 1996 ; 9 : 62 -67. 14 Abraham EC, Perry RE, Stallings M. Application of affinity chromatography for separation and quantification of glycosylated hemoglobins. J Lab Clin Med 1983 ; 102 : 187 -197. PubMed 15 Garlick RL, Mazer JS, Higgins PJ, Bunn HF. Characterization of glycosylated hemoglobins: relevance to monitoring of diabetic control and analysis of other proteins. J Clin Invest 1983 ; 71 : 1062 -1072. Crossref Search ADS PubMed 16 Fluckinger R, Woodtli T. Effect of temperature on quantifying glycated (glycosylated) hemoglobin by cation exchange chromatography. Clin Chem 1985 ; 31 : 114 -117. Crossref Search ADS PubMed 17 Cohen MP, Witt J, Wu VY. Purified hemoglobin preparations in the evaluation of HbA1c determination by ion exchange chromatography. Ann Clin Biochem 1993 ; 30 : 265 -271. Crossref Search ADS PubMed 18 Turpeinen U, Karjaleinen U, Stenman U. Three assays for glycohemoglobin compared. Clin Chem 1995 ; 41 : 191 -195. Crossref Search ADS PubMed 19 Koskinen LK. Specificity of hemoglobin A1c measurement by cation exchange liquid chromatography. Evaluation of a Mono S column method. Clin Chim Acta 1996 ; 253 : 159 -169. Crossref Search ADS PubMed 20 Eckerbom S, Bergqvist Y, Jeppsson J-O. Improved method for analysis of glycated haemoglobin by ion exchange chromatography. Ann Clin Biochem 1994 ; 31 : 1 -6. Crossref Search ADS PubMed 21 Jeppsson J-O, Jerntorp P, Sundkvist G, Englund H, Nylund U. Measurements of hemoglobin A1c by a new liquid-chromatographic assay: methodology, clinical utility and relation to glucose tolerance evaluated. Clin Chem 1986 ; 32 : 1867 -1872. Crossref Search ADS PubMed 22 Miedema K, Goldstein D, Goodall I, Hoelzel W, Jeppsson JO, Little R, Shima K. IFCC Working Group on Standardization of HbA1c [Abstract]. Martin SM Halloren SP eds. Proceedings of the XVI Congress of Clinical Chemistry 1996 : 39 Pigott Printers Cambridge, UK. . 23 Finke A, Hoelzel W, Jeppsson JO, Kobold U, Miedema K, Penders TJ, Weykamp C. Development of a candidate primary reference material for the international standardization of HbA1c [Abstract]. Martin SM Halloren SP eds. Proceedings of the XVI Congress of Clinical Chemistry 1996 : 374 Pigott Printers Cambridge, UK. . 24 Stöckl D, Franzini C, Kratochvila J, Middle J, Ricos C, Siekmann L, Thienpont LM. Analytical specifications of Reference Methods. Compilation and critical discussion (from members of the European EQA-Organizers Working Group B) [Review]. Eur J Clin Chem Clin Biochem 1996 ; 34 : 319 -337. PubMed 25 Fenn JB, Mann M, Meng C, Wong SF, Whitehouse CM. Electrospray ionization for mass spectrometry of large biomolecules. Science 1989 ; 246 : 64 -71. Crossref Search ADS PubMed 26 . International Committee for Standardization in Haematology. Recommendations for hemoglobinometry in human blood. Br J Haematol 1967 ; 13 : 71 -75. Crossref Search ADS PubMed 27 Jeppsson J-O, Källman I, Lindgren G, Fägerstam LG. Hb-Linköping (β 36 Pro-Thr). A new hemoglobin mutant characterized by reversed phase high performance liquid chromatography. J Chromatogr 1984 ; 297 : 31 -36. Crossref Search ADS PubMed © 1997 The American Association for Clinical Chemistry This article is published and distributed under the terms of the Oxford University Press, Standard Journals Publication Model (https://academic.oup.com/journals/pages/open_access/funder_policies/chorus/standard_publication_model) TI - Candidate reference methods for hemoglobin A1c based on peptide mapping JF - Clinical Chemistry DO - 10.1093/clinchem/43.10.1944 DA - 1997-10-01 UR - https://www.deepdyve.com/lp/oxford-university-press/candidate-reference-methods-for-hemoglobin-a1c-based-on-peptide-kgJbf2ht79 SP - 1944 EP - 1951 VL - 43 IS - 10 DP - DeepDyve ER -