TY - JOUR AU - Abe,, Yoshihisa AB - Abstract In a previous study, we reported on a novel (prototype) real-time patient dosimeter with non-toxic phosphor sensors. In this study, we developed new types of sensors that were smaller than in the previous prototype, and clarified the clinical feasibility of our newly proposed dosimeter. Patient dose measurements obtained with the newly proposed real-time dosimeter were compared with measurements obtained using a calibrated radiophotoluminescence glass reference dosimeter (RPLD). The reference dosimeters were set at almost the same positions as the new real-time dosimeter sensors. We found excellent correlations between the reference RPLD measurements and those obtained using our new real-time dosimeter (r2 = 0.967). However, the new type of dosimeter was found to underestimate radiation skin dose measurements when compared with an RPLD. The most probable reason for this was the size reduction in the phosphor sensor of the new type of dosimeter. We believe that, as a result of reducing the phosphor sensor size, the backscattered X-ray irradiation was underestimated. However, the new dosimeter can accurately determine the absorbed dose by correcting the measured value with calibration factors. The calibration factor for the new type dosimeter was determined (by linear regression) to be ~1.15. New real-time patient dosimeter design would be an effective tool for the real-time measurement of patient skin doses during interventional radiology treatments. INTRODUCTION Although numerous patients have greatly benefited from cardiac interventional radiology (IR), radiation-induced skin injuries have been reported in several cardiac IR procedures [1–12]. These deterministic effects arise because these complicated procedures tend to require an extended fluoroscopic exposure time and repeated digital angiography (DA) with various angles of X-ray projection. In addition, these procedures often need to be performed repeatedly. Thus, in order to increase the efficacy and safety of cardiac IR procedures, one must therefore reduce the patient’s exposure to excessive X-ray treatments. In addition, the real-time monitoring of patient radiation doses is essential, and in particular, the peak radiation skin dose (PSD) should be monitored in real-time to avoid radiation-induced skin injury [1, 13–25]. Recently, the International Commission on Radiological Protection (ICRP) emphasized the necessity of using diagnostic reference levels (DRLs) to optimize patient doses and recommended that, during actual IR procedures, the maximum cumulative absorbed dose in the skin should be monitored in real-time [25]. However, although the real-time monitoring of patient doses is important, there are few feasible real-time dosimeters currently available for IR patients [26]. In a previous study, we reported on a novel (prototype) real-time patient dosimeter with four non-toxic phosphor sensors suitable for red-emission phosphor in the medical X-ray range [27–29] (Fig. 1). We conducted a feasibility study in a clinical setting to investigate the use of the prototype dosimeter during a number of different procedures [coronary angiography (CAG), radiofrequency catheter ablation (RFCA), and percutaneous coronary intervention (PCI)]. Fig. 1. View largeDownload slide New type of real-time patient dosimeter. Fig. 1. View largeDownload slide New type of real-time patient dosimeter. In this study, we developed new types of sensors that were smaller than in the previous prototype (Fig. 2) and measured the patient skin absorbed dose in RFCA treatments. RFCA is a standard treatment modality for tachyarrhythmia and is performed to disconnect the pathway involved in the abnormal rhythm. Significantly, RFCA is an effective procedure for adults as well as children; however, RFCA tends to be a complex procedure, which increases the fluoroscopic time and can deliver high radiation doses to the patient. Moreover, the patient’s applied radiation dose depends upon the type of arrhythmia they exhibit. Fig. 2. View largeDownload slide Non-toxic phosphor sensor (Y2O2S: Eu, Sm). Fig. 2. View largeDownload slide Non-toxic phosphor sensor (Y2O2S: Eu, Sm). MATERIALS AND METHODS Patients This study included 40 RFCA procedures and was conducted at the Research Institute for Brain and Blood Vessels – Akita (Akita, Japan). The patients were selected at random and included 26 males and 14 females (Table 1). This was a single-institution study and was approved by the local Committee on Human Research. Table 1. Summary of the patient characteristics (mean ± standard deviations where applicable) for radiofrequency catheter ablation (n = 40) Age (years) Height (cm) Weight (kg) Male/Female Fluoroscopy time (min) Cumulative air kerma (mGy) Kerma area product (Gy · cm2) 59.5 ± 14.4 164.7 ± 11.4 61 ± 12.9 26/14 25.3 ± 15.5 421.6 ± 428.8 59.9 ± 50.3 Age (years) Height (cm) Weight (kg) Male/Female Fluoroscopy time (min) Cumulative air kerma (mGy) Kerma area product (Gy · cm2) 59.5 ± 14.4 164.7 ± 11.4 61 ± 12.9 26/14 25.3 ± 15.5 421.6 ± 428.8 59.9 ± 50.3 Table 1. Summary of the patient characteristics (mean ± standard deviations where applicable) for radiofrequency catheter ablation (n = 40) Age (years) Height (cm) Weight (kg) Male/Female Fluoroscopy time (min) Cumulative air kerma (mGy) Kerma area product (Gy · cm2) 59.5 ± 14.4 164.7 ± 11.4 61 ± 12.9 26/14 25.3 ± 15.5 421.6 ± 428.8 59.9 ± 50.3 Age (years) Height (cm) Weight (kg) Male/Female Fluoroscopy time (min) Cumulative air kerma (mGy) Kerma area product (Gy · cm2) 59.5 ± 14.4 164.7 ± 11.4 61 ± 12.9 26/14 25.3 ± 15.5 421.6 ± 428.8 59.9 ± 50.3 This study was conducted in accordance with the Declaration of Helsinki. All protocols were approved by Research Institute for Brain and Blood Vessels – Akita (Akita, Japan). Informed consent was obtained from patients before commencing procedures. X-ray equipment The main angiography X-ray unit used in this study was a digital cine single-plane system (Artis Zee FA; Siemens Healthcare, Germany) with a 32-cm mode flat-panel detector. Digital cine acquisition was performed at 7.5 frames/s, and pulsed fluoroscopy was performed at 7.5 pulses/s. Four standard tube angulations were used in our clinical RFCA setting: right anterior oblique (RAO) 60° view, RAO 30° view, antero–posterior (AP) view, and left anterior oblique (LAO) 50° view. Dosimetry Our new smaller real-time dosimeter comprised photoluminescence sensors (non-toxic phosphor, maximum of four sensors), an optical fiber cable, a photodiode, and a digital display that included the power supply [28]. The new sensor used the same Y2O2S: Eu, Sm phosphor material as was used for the prototype, since we note that Y2O2S: Eu, Sm is non-toxic and exhibits relatively high sensitivity. The photoluminescence sensor size of prototype was a square of 2.5 × 2.5 mm and the photoluminescence sensor size of new type was a circle of radius 0.7 mm. The sensor area of new type was reduced to 1/4. Patient dose measurements obtained with the newly proposed real-time dosimeter were compared with measurements obtained using a calibrated radiophotoluminescence glass reference dosimeter (RPLD) [30–32]. The measurement/readout system used for the RPLD (GD-302 M sensor; Asahi Techno Glass, Tokyo, Japan) was the DoseAce FGD-1000 unit (Chiyoda Technol Corp., Tokyo, Japan). The calibration factor for the RPLD used in this study was determined to be 0.348, based on a calibration experiment conducted in air using a thimble-type 6-ml ion-chamber (model-9015; Radcal Corp., Monrovia, CA, USA; traceable from the national standard exposure dose). We previously reported on a modified dosimetry gown that can measure the entrance skin dose during PCI using multiple RPLDs [31]. From the study, we created the entrance skin dose distribution map of patients, and we could comprehend the irradiation field of each angle. The measured radiation skin dose was examined using the four sensors of the new real-time dosimeter, arbitrarily positioned at the fields of the RFCA working angle. The reference dosimeters were set at almost the same positions (within 1 cm) of the new real-time dosimeter sensors. The RPLD and the new real time dosimeter were responsive to backscattered radiation, and the measurement doses were entrance surface dose (air kerma). In our clinic, the entrance surface dose was transformed to skin absorbed dose based on the ‘Tables of X-Ray Mass Attenuation Coefficients’ by S. M. Seltzer and J. H. Hubbell [33]. We also interviewed physicians regarding the visibility (i.e. whether the view was impeded) of the radiographic images in which the new dosimeter sensors were visible. Finally, the relationship between the reference and the new type of dosimeter measurements was analyzed using linear regression. RESULTS Figure 3 shows the correlations between the RPLD measurements and the real-time dosimeter measurements. The prototype sensor and the new type of sensor have excellent correlation coefficients between the RPLD measurements. From the linear regression fit, y = 1.15x + 3.62 (r2 = 0.968), the new type of dosimeter is seen to underestimate the dose by 15% with respected to the calibrated reference RPLD dosimeter. Fig. 3. View largeDownload slide Relationship between the reference dosimeter (radiophotoluminescence glass dosimeter; RPLD) measurements of the patient radiation dose and those of the new type of dosimeter in a clinical setting. Fig. 3. View largeDownload slide Relationship between the reference dosimeter (radiophotoluminescence glass dosimeter; RPLD) measurements of the patient radiation dose and those of the new type of dosimeter in a clinical setting. Figure 4 shows the Pearson correlation test revealed statistically positive correlations between the measurements of each irradiation field obtained using the RPLDs and those measurements obtained using the new type of sensor, RAO60 (r = 0.997, P < .01), RAO30 (r = 0.980, P < .01), P-A (r = 0.966, P < .01) and LAO50 (r = 0.986, P < .01), respectively. Thus, we concluded that new real-time dosimeter had no angular dependence. Fig. 4. View largeDownload slide Angular dependence of new real-time dosimeter. RAO = right anterior oblique, LAO = left anterior oblique. Fig. 4. View largeDownload slide Angular dependence of new real-time dosimeter. RAO = right anterior oblique, LAO = left anterior oblique. Figures 5 and 6 show an angiogram taken during a patient’s dose measurement in an RFCA procedure. Although the prototype dosimeter sensor (Fig. 5) is quite noticeable, our new smaller dosimeter sensor (Fig. 6) is only slightly noticeable. In both cases, the dosimeter cable is invisible in the angiographic image, in addition to the reference RPLDs. We note that the number of complaints received from physicians and other medical staff regarding obstruction caused by the new type of sensors during the procedure was less than the number received for the original prototype dosimeter sensor. Fig. 5. View largeDownload slide Angiogram taken during a patient’s dose measurement in an RFCA procedure. Arrows indicate the locations of the prototype dosimeter sensors. The size of the photoluminescence sensor was a square of 2.5 × 2.5 mm. RFCA = radiofrequency catheter ablation, RAO = right anterior oblique, LAO = left anterior oblique. Fig. 5. View largeDownload slide Angiogram taken during a patient’s dose measurement in an RFCA procedure. Arrows indicate the locations of the prototype dosimeter sensors. The size of the photoluminescence sensor was a square of 2.5 × 2.5 mm. RFCA = radiofrequency catheter ablation, RAO = right anterior oblique, LAO = left anterior oblique. Fig. 6. View largeDownload slide Angiogram taken during a patient’s dose measurement in an RFCA procedure. Arrows indicate the locations of the new type of dosimeter sensors. The size of the photoluminescence sensor was a circle of radius 0.7 mm. RFCA = radiofrequency catheter ablation, RAO = right anterior oblique, LAO = left anterior oblique. Fig. 6. View largeDownload slide Angiogram taken during a patient’s dose measurement in an RFCA procedure. Arrows indicate the locations of the new type of dosimeter sensors. The size of the photoluminescence sensor was a circle of radius 0.7 mm. RFCA = radiofrequency catheter ablation, RAO = right anterior oblique, LAO = left anterior oblique. DISCUSSION According to ICRP Publication 105, DRLs for IR are used to promote the management of patient doses with a view to avoiding stochastic effects. In addition, the document recommends that the PSD be measured during actual IR procedures to avoid deterministic effects. This study was performed to clarify the clinical feasibility of our new type of patient dosimeter. We found an excellent correlation between the patient skin dose as measured by a reference RPLD and as measured by our new type of dosimeter in a clinical setting (r2 = 0.968). Although both our original prototype sensor and the new sensor are slightly visible on clinical images, the new sensor is significantly smaller than the prototype sensor. Therefore, the new sensor dose not impact the coronary artery images significantly. We conclude that the newly proposed dosimeter is also feasible for clinical use in CAG and PCI procedures. However, the new type of dosimeter was found to underestimate radiation skin dose measurements when compared with a calibrated reference RPLD dosimeter. The most probable reason for this was a reduction in the phosphor sensor size of the new type of dosimeter. As a result of reducing the phosphor sensor size, we expect that the backscatter X-ray was underestimated, and that the collection efficiency between the phosphor sensor and the optical fiber cable decreased. However, the new dosimeter can measure an accurate (i.e. in accordance with the reference RPLD dosimeter) absorbed dose by correcting the value with calibration factors. The calibration factor of the new type dosimeter was determined by linear regression to be ~1.15. Some differences were also found between the reference RPLD measurements and the new type of dosimeter measurements. One probable reason for these differences is that the new sensor type is angle-dependent. Although both sets of sensors were located at approximately the same positions, there was a small distance (~1 cm) between the sensor positions. This means that, in some cases, one of the sensors (i.e. the RPLD or our new dosimeter) may not have been located within the primary X-ray beam for some of the projections used during each procedure. In conclusion, we found excellent correlations between the reference radiation skin dose measurements and those obtained using our newly proposed real-time dosimeter (r2 = 0.968). Therefore, our new device can be used to accurately measure radiation skin doses at the specified measuring points in a clinical setting using a calibration factor of 1.15. Moreover, the sensor and cable of the proposed dosimeter do not interfere with the IR procedure. As a result, we believe that our new photoluminescence sensor dosimeter design would be an effective tool for the real-time measurement of patient skin doses during IR treatments. ACKNOWLEDGEMENTS We thank Dr Takashi Moritake, University of Occupational and Environmental Health, Japan, for his invaluable assistance. We thank Dr Son Lue from the Institute of Clinical Medicine, Graduate School of Comprehensive Human Sciences, University of Tsukuba, for helpful advice regarding the technical analysis. We also thank Mr Fumiaki Sasaki and Mr Kazunori Matsumoto from the Department of Radiology and Nuclear Medicine, Research Institute for Brain and Blood Vessels—Akita, for their invaluable assistance. CONFLICT OF INTEREST The authors declare that there is no financial conflict of interest to be disclosed. FUNDING This work was partially supported by a Grant-in-Aid for Scientific Research [No. 16K09029] from the Japan Society for the Promotion of Science. This study was supported in part by the Radiation Safety Research Promotion Fund, Nuclear Regulation Authority, Japan. REFERENCES 1 Balter S , Miller DL . Patient skin reactions from interventional fluoroscopy procedures . AJR Am J Roentgenol 2014 ; 202 : W335 – 42 . Google Scholar Crossref Search ADS PubMed 2 International Commission on Radiological Protection . ICRP publication 85: avoidance of radiation injuries from medical interventional procedures . Ann ICRP 2000 ; 30 : 7 – 67 . 3 Chida K , Inaba Y , Saito H et al. Radiation dose of interventional radiology system using a flat-panel detector . AJR Am J Roentgenol 2009 ; 193 : 1680 – 5 . Google Scholar Crossref Search ADS PubMed 4 Chida K , Ohno T , Kakizaki S et al. Radiation dose to the pediatric cardiac catheterization and intervention patient . AJR Am J Roentgenol 2010 ; 195 : 1175 – 9 . Google Scholar Crossref Search ADS PubMed 5 Chida K , Takahashi T , Ito D et al. Clarifying and visualizing sources of staff-received scattered radiation in interventional procedures . AJR Am J Roentgenol 2011 ; 197 : W900 – 3 . Google Scholar Crossref Search ADS PubMed 6 Dauer LT , Thornton RH , Hay JL et al. Fears, feelings, and facts: interactively communicating benefits and risks of medical radiation with patients . AJR Am J Roentgenol 2011 ; 196 : 756 – 61 . Google Scholar Crossref Search ADS PubMed 7 Chida K , Kaga Y , Haga Y et al. Occupational dose in interventional radiology procedures . AJR Am J Roentgenol 2013 ; 200 : 138 – 41 . Google Scholar Crossref Search ADS PubMed 8 Kato M , Chida K , Sato T et al. The necessity of follow-up for radiation skin injuries in patients after percutaneous coronary interventions: radiation skin injuries will often be overlooked clinically . Acta Radiol 2012 ; 53 : 1040 – 4 . Google Scholar Crossref Search ADS PubMed 9 International Commission on Radiological Protection . ICRP publication 120. Radiological protection in cardiology . Ann ICRP 2013 ; 42 : 1 – 125 . 10 Chida K , Saito H , Zuguchi M et al. Does digital acquisition reduce patients’ skin dose in cardiac interventional procedures? An experimental study . AJR Am J Roentgenol 2004 ; 183 : 1111 – 4 . Google Scholar Crossref Search ADS PubMed 11 Tsapaki V , Ahmed NA , AlSuwaidi JS et al. Radiation exposure to patients during interventional procedures in 20 countries: initial IAEA project results . AJR Am J Roentgenol 2009 ; 193 : 559 – 69 . Google Scholar Crossref Search ADS PubMed 12 Santos WS , Neves LP , Perini AP et al. Exposures in interventional radiology using Monte Carlo simulation coupled with virtual anthropomorphic phantoms . Phys Med 2015 ; 31 : 929 – 33 . Google Scholar Crossref Search ADS PubMed 13 Chida K , Saito H , Otani H et al. Relationship between fluoroscopic time, dose–area product, body weight, and maximum radiation skin dose in cardiac interventional procedures . AJR Am J Roentgenol 2006 ; 186 : 774 – 8 . Google Scholar Crossref Search ADS PubMed 14 Morishima Y , Chida K , Meguro T . Effectiveness of additional lead shielding to protect staff from scattering radiation during endoscopic retrograde cholangiopancreatography procedures . J Radiat Res 2018 ; 59 : 225 – 32 . Google Scholar Crossref Search ADS PubMed 15 Chida K , Kagaya Y , Saito H et al. Total entrance skin dose: an effective indicator of maximum radiation dose to the skin during percutaneous coronary intervention . AJR Am J Roentgenol 2007 ; 189 : W224 – 7 . Google Scholar Crossref Search ADS PubMed 16 Chida K , Inaba Y , Masuyama H et al. Evaluating the performance of a MOSFET dosimeter at diagnostic X-ray energies for interventional radiology . Radiol Phys Technol 2009 ; 2 : 58 – 61 . Google Scholar Crossref Search ADS PubMed 17 Chida K , Kato M , Kagaya Y et al. Radiation dose and radiation protection for patients and physicians during interventional procedure . J Radiat Res 2010 ; 51 : 97 – 105 . Google Scholar Crossref Search ADS PubMed 18 Struelens L , Bacher K , Bosmans H et al. Establishment of trigger levels to steer the follow-up of radiation effects in patients undergoing fluoroscopically-guided interventional procedures in Belgium . Phys Med 2014 ; 30 : 934 – 40 . Google Scholar Crossref Search ADS PubMed 19 Kato M , Chida K , Sato T et al. Evaluating the maximum patient radiation dose in cardiac interventional procedures . Radiat Prot Dosimetry 2011 ; 143 : 69 – 73 . Google Scholar Crossref Search ADS PubMed 20 Miller DL , Balter S , Dixon RG et al. . Society of Interventional Radiology Standards of Practice Committee . Quality improvement guidelines for recording patient radiation dose in the medical record for fluoroscopically guided procedures . J Vasc Interv Radiol 2012 ; 23 : 11 – 8 . Google Scholar Crossref Search ADS PubMed 21 Sun L , Mizuno Y , Iwamoto M et al. Direct measurement of a patient’s entrance skin dose during pediatric cardiac catheterization . J Radiat Res 2014 ; 55 : 1122 – 30 . Google Scholar Crossref Search ADS PubMed 22 Inaba Y , Chida K , Kobayashi R et al. A cross-sectional study of the radiation dose and image quality of X-ray equipment used in IVR . J Appl Clin Med Phys 2016 ; 17 : 391 – 401 . Google Scholar Crossref Search ADS PubMed 23 Farah J , Cuttat M , Hadid L et al. Patient dosimetry in interventional radiology: uncertainties associated to skin dose measurement and exposure correlation to online dose indicators . Phys Med 2015 ; 31 : e46 – 7 . Google Scholar Crossref Search ADS 24 Takata T , Kotoku J , Maejima H et al. Fast skin dose estimation system for interventional radiology . J Radiat Res 2018 ; 59 : 233 – 9 . Google Scholar Crossref Search ADS PubMed 25 International Commission on Radiological Protection . ICRP publication 105. Radiological protection in medicine . Ann ICRP 2007 ; 37 : 1 – 63 . 26 Chida K , Kagaya Y , Saito H et al. Evaluation of patient radiation dose during cardiac interventional procedures: what is the most effective method? Acta Radiol 2009 ; 50 : 474 – 81 . Google Scholar Crossref Search ADS PubMed 27 Nakamura M , Chida K , Zuguchi M . Red emission phosphor for real-time skin dosimeter for fluoroscopy and interventional radiology . Med Phys 2014 ; 41 : 101913 . Google Scholar Crossref Search ADS PubMed 28 Nakamura M , Chida K , Zuguchi M . Novel dosimeter using a nontoxic phosphor for real-time monitoring of patient radiation dose in interventional radiology . AJR Am J Roentgenol 2015 ; 205 : W202 – 6 . Google Scholar Crossref Search ADS PubMed 29 Chida K , Kato M , Inaba Y et al. Real-time patient radiation dosimeter for use in interventional radiology . Phys Med 2016 ; 32 : 1475 – 8 Google Scholar Crossref Search ADS PubMed 30 Moritake T , Hayakawa M , Matsumaru Y et al. Precise mapping system of entrance skin dose during endovascular embolization for cerebral aneurysm . Radiat Meas 2011 ; 46 : 2103 – 6 . Google Scholar Crossref Search ADS 31 Kato M , Chida K , Moritake T et al. Fundamental study on the characteristics of a radiophotoluminescence glass dosemeter with no energy compensation filter for measuring patient entrance doses in cardiac interventional procedures . Radiat Prot Dosimetry 2014 ; 162 : 224 – 9 . Google Scholar Crossref Search ADS PubMed 32 Kato M , Chida K , Moritake T et al. Direct dose measurement on patient during percutaneous coronary intervention procedures using radiophotoluminescence glass dosimeters . Radiat Prot Dosimetry 2017 ; 175 : 31 – 7 . Google Scholar PubMed 33 National Institute of Standards and Technology (NIST) (Hubbell JH, Seltzer SM). NIST Standard Reference Database 126. https://www.nist.gov/pml/x-ray-mass-attenuation-coefficients?iframe=true&width=100%25&height=100%25 (24 September 2018, date last accessed). © The Author(s) 2019. Published by Oxford University Press on behalf of The Japan Radiation Research Society and Japanese Society for Radiation Oncology. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited. TI - New real-time patient radiation dosimeter for use in radiofrequency catheter ablation JO - Journal of Radiation Research DO - 10.1093/jrr/rry110 DA - 2019-03-01 UR - https://www.deepdyve.com/lp/oxford-university-press/new-real-time-patient-radiation-dosimeter-for-use-in-radiofrequency-io0v0VZD52 SP - 215 VL - 60 IS - 2 DP - DeepDyve ER -