TY - JOUR AU - Tikhomandritskaya, V. AB - Let us now, on the basis of the same data as we used for the previous example, work out an example. It is obvious that A 1 = 0.2,andA2=200O. Then AA t = 0.2X 7 = 1.4 and BA 2 = 200 x 0.01 = 2. The difference of these two quantities is -0.6. It is obvious that [-0.6] < 16. Hence, the error of the attenuator with respect to its loss does not exceed the tolerance. It should be noted that the attenuators should not only be tested on direct but also on al- ternating current, especially after repairs or when attenuators of an unknown design are being tested. It is obvious that the above reasoning can be extended to the checking of attenuators by means of an alternating current, when it will be necessary to measure impedances Z o and Z s with an error of 0.1 - 0.2 %. LITERATURE CITED 1. N.N. Solov'ev, Foundations of Measurement Techniques in Wire Communications [in Russian]. Moscow- Leningrad, 1959, v. rrl. V. A. Tikhomandritskaya Translated from hmeritel'naya Tekhnika, No. 7o pp. 41-42, July, 1962 Coaxial connectors with a small reflection coefficient are required for precision instruments of voltage stand- TI - Coaxial plug-type connector with a low reflection coefficient JF - Measurement Techniques DO - 10.1007/BF00989033 DA - 2004-12-30 UR - https://www.deepdyve.com/lp/springer-journals/coaxial-plug-type-connector-with-a-low-reflection-coefficient-dIgwltTLos SP - 586 EP - 588 VL - 5 IS - 7 DP - DeepDyve ER -