TY - JOUR AU - Crawford,, Gregory AB - Abstract Background: Anemia is an underdiagnosed, significant public health concern afflicting >2 billion people worldwide. The detrimental effects of tissue oxygen deficiency on the cardiovascular system and concurrent appearance of anemia with numerous high-risk disorders highlight the importance of clinical screening. Currently there is no universally accepted, clinically applicable, noninvasive hemoglobin/hematocrit screening tool. The need for such a device has prompted an investigation into a breadth of techniques. Methods: A synopsis of the literature and current directions of research in noninvasive total hemoglobin measurement was collected. Contributions highlighted in this review are limited to those studies conducted with a clinical aspect, and most include in vivo patient studies. Results: The review of potential techniques presented here includes optoacoustic spectroscopy, spectrophotometric imaging, diffuse reflectance spectroscopy, transcutaneous illumination, electrical admittance plethysmography, and photoplethysmography. The technological performance, relative benefits of each approach, potential instrumentation design considerations, and future directions are discussed in each subcategory. Conclusions: Many techniques reviewed here have shown excellent accuracy, sensitivity, and specificity in measuring hemoglobin/hematocrit, thus in the near future a new clinically viable tool for noninvasive hemoglobin/hematocrit monitoring will likely be widely used for patient care. Limiting factors in clinical adoption will likely involve technology integration into the current standard of care in each field routinely dealing with anemia. Although anemia is often perceived by the general population as a minor medical condition, it is a significant ailment that healthcare professionals have recognized as negatively affecting mortality and morbidity (1)(2). According to the WHO, anemia is the largest global illness adversely affecting mortality and worker capacity. The US Department of Health and Human Services has deemed anemia a significant public health concern, and the National Anemia Action Council has reported the estimate that although anemia has been diagnosed in 3.5 million people in the US, millions more cases go undiagnosed. In developing countries where nutritional inadequacies and infectious disease are more prevalent, the situation effects of anemia are thought to be amplified, severely hindering children from reaching their full genetically determined potential (3)(4). The WHO estimates that as many as 2 billion people worldwide are afflicted with anemia (5). Anemia is the lack of healthy red blood cells (RBC)1 circulating in the vascular system. Medically, anemia is defined by the WHO as a hemoglobin (Hgb) concentration below 12 g/dL for females and below 13 g/dL for males (5)(6). Even in its mild forms, anemia can influence physical function through fatigue and weakness; it also decreases myocardial function, causes peripheral arterial vasodilation, and activates the sympathetic and renin-angiotensin-aldosterone system. These effects influence the progression of diseases such as cardiac and renal failure (7)(8). In addition, anemia is associated with a myriad of other diseases. For example, anemia affects at least 33% of all patients with cancer, an estimated 65%–95% of all patients with HIV/AIDS, and 70% of all patients with rheumatoid arthritis (9). To screen for anemia, physicians currently perform a visual inspection of the palpebral conjunctiva, conduct a complete blood count (CBC) test, spin a hematocrit (Hct), or use a small blood volume Hgb meter. Visual inspection of the conjunctiva by a physician is, at best, 70% accurate independent of the physician’s experience and training; it has also been shown that physicians today are less accurate than those of the past (10). The CBC test is very accurate but is invasive and painful to the patient, associated with a significant cost, time-consuming in waiting for laboratory analysis, and often not part of a routine physical exam. Spinning an Hct or using a small blood volume Hgb monitor provides a mode of quick anemia detection, but it is painful to the patient while creating biohazards and the need for associated disposal procedures. A device that can quickly and accurately measure Hgb has many healthcare applications, such as in physical examinations, emergency departments, prehospital care providers, medical specialists, in situ measurement of bleeding during surgery for determining transfusion triggers, and in home healthcare for the chronically ill and aging population. A noninvasive portable tool for anemia screening can make a difference at emergency department triage points, in field treatment centers during large-scale disasters (11), and in austere environments where rapid triage and treatment decisions are based on few data points. The use of devices that require blood may not be suitable in these settings. We present a review of noninvasive techniques to measure Hct/Hgb, either of which provides a quantification of anemia. It is important to note that although correlation coefficients are given for most multipatient trials, the correlation methods and reference techniques vary with each study and these numbers are meant only to be a rough guide. conductance methods In 1980, Yamakoshi et al. reported on the potential of monitoring Hct using electrical admittance plethysmography (12). During pulsation, the change in blood volume in the finger translates to a change in electrical admittance (conductivity) of the finger. Submerging the finger in an electrolyte solution whose admittance is equal to that of the finger compensates for pulsatile variation in conductivity, after which the conductivity of the electrolyte solution can be related to the conductivity of arterial vessels and then correlated to the Hct. This technique was reported with a correlation to linear regression line of r = 0.98 (n = 16), determined by calibration established from in vitro blood conductance measurements and compared to a capillary-centrifuged reference sample. This group further disclosed a more simplistic and potentially viable clinical method to implement this technique using an electrical admittance finger cuff (13). Electrodes are placed in the interior of an annular cuff, which is then filled with an electrolyte solution. With the finger inserted through the cuff, electrical admittance measurements are taken and related to the electrolyte solution conductance in a similar fashion to the submersion technique, translating to a cleaner and more portable device. Although the accuracy reported for the conductance method is promising, albeit in a limited patient population, little recent work has been disclosed on this technique. Difficulties associated with probing conductivity include variations with temperature, intra- and extracellular water/ion content, and other blood constituents. Because conductance monitors only a single value, rejection sodium ion concentration variation becomes a limiting factor, particularly in a noninvasive setting where varying whole blood ion content cannot be measured and thus cannot be normalized, hindering the potential overall efficacy of this technique. This topic is not explicitly discussed in this contribution. imaging and spectrophotometric imaging Because Hgb (both oxy- and deoxy- forms) is one of the primary chromophores in blood, an assortment of optical techniques has been evaluated in noninvasive Hgb screening devices. One subcategory of device includes combination near-infrared (NIR) vascular imagers/spectrophotometers that can target blood vessels and subsequently estimate Hgb concentration based on the absorption characteristics of individual vessels. Kanashima et al. reviewed the performance of one such device, an NIR-imaging based noninvasive Hgb monitor (Astrim™ Sysmex) (14)(15). Several NIR wavelength bands are transmitted through the fingertip and used to image blood vessels and subsequently calculate Hgb concentration. Explicitly, photometric absorbance measurements from the vascular portions of the image are coupled with an estimation of the optical pathlength using the imaged vascular diameter (and assuming circular vessel cross sections). This group reported diagnostic sensitivity and specificity of 78.3% and 69%, respectively, for clinically defined anemia, with a correlation to linear regression line of r = 0.53 for anemic patients (n = 174), r = 0.34 (n = 135) for patients with normal Hgb levels, and r = 0.59 (n = 309) for all patients, using predefined device calibrations and comparing results to an automated hematology analyzer. Although this group notes the device is poorly suited for Hgb determination from a single measurement owing to variability in efficacy with finger position, its precision lends to its suitability for tracking of Hgb levels in perioperative anemia, gastrointestinal bleeding, and other situations requiring dynamic blood monitoring. As a more accurate solution, numerous groups have examined spectrophotometric imaging at other physiological access points. Nadeau and Groner have disclosed an alternative method of image collection to probe microvascular networks and measure Hgb noninvasively using polarization modulation (16). Orthogonal polarization spectral (OPS) imaging was implemented in a commercial instrument (Hemoscan, Cytometrics) to analyze the vascular network of the sublingual mucosa. Illuminating with a polarized source at a discrete spectral band of high Hgb absorbance (approximately 550 nm), a crossed polarizer is placed at the detection focal plane array that transmits only light that has been depolarized through scattering from deep penetration into tissue. Hgb concentration is determined on the basis of a computational estimation of the vascular network density from collected images and the intensity of reflectance signal from each vessel area. This group has reported a linear regression correlation of r = 0.93 (n = 71) for predefined instrument calibrations compared to an automated hematology analyzer. Beyond Hgb measurement, this mode of imaging also provides insight into underlying vascular disorders. Genzel-Boroviczeny et al. present a similar imaging technique using OPS to measure Hgb noninvasively in neonates (17), and a report by Winkelman discusses the technology and further clinical implications/viability of this microcirculation imaging technique (18). Rice et al. have reported on the potential of standardized retinal imaging as an indicator for Hgb, serum bilirubin, and glucose (19). The retina is illuminated through the pupil, preferably at isosbestic points for oxy- and deoxyhemoglobin, using several discrete visible/NIR light bands while the reflected light from vessels overlying the optic disk is analyzed. This technique yields a cross-validation correlation of r = 0.89 (n = 24) based on calibrated reflected intensities compared with a small volume in vitro Hgb analyzer (HemoCue B, HemoCue AB). Winkelman et al. describe a similar image analysis technique for Hgb concentration estimation that focuses on the bulbar conjunctiva capillaries rather than the retinal vessels (20). This method analyzes high-magnification images of the capillary vessels in the mucosal surface to estimate blood cell parameters, including Hgb. Iftimia et al. present an alternative imaging mode of Hct determination using spectral-domain low coherence interferometry (21)(22), an alternative configuration of optical coherence tomography for retinal imaging that enables faster data collection. In this configuration, the penetration depth of optical signals through a blood vessel is dependent on the amount of scatter, which in turn is correlated to Hct. Fig. 1 shows an image of retinal vessels obtained with tracking laser-scanning ophthalmoscopy and the correlated optical coherence tomography depth profiles. To maintain collection from a single retinal vessel in the presence of rapid eye movements, an active eye tracking system is integrated to maintain image fixation on a single blood vessel (21). Faubert et al. report a similar technique using reflectance measurements from the retinal vessels to correlate to Hgb concentration (23)(24). Figure 1. Open in new tabDownload slide Retinal imaging for Hct determination. (A), Scanning image of retinal vessels using a laser tracking imager. (B), OCT image along the horizontal line in (A). (C), OCT image along the vertical line in (A). Hct is correlated based on depth profiles of individual blood vessels. Reproduced with permission from (21). Figure 1. Open in new tabDownload slide Retinal imaging for Hct determination. (A), Scanning image of retinal vessels using a laser tracking imager. (B), OCT image along the horizontal line in (A). (C), OCT image along the vertical line in (A). Hct is correlated based on depth profiles of individual blood vessels. Reproduced with permission from (21). The physiological structures discussed in the above studies, such as the retina, conjunctiva, and sublingual mucosa, are mucosal surfaces, and as such are excellent locations to noninvasively monitor blood analytes compared to thick tissue regions such as the finger or forearm. In these regions (1), the transparency of overlaying mucous membrane allows for clear observation and imaging of underlying vessels (2), and low melanin concentrations in these areas leads to high uniformity between patients with different skin pigments (3). Analysis of capillary beds in a mucosal surface yields stable optical pathlengths during pulsation and thus does not require acquisition during predetermined timeframes of systole and diastole. Although capillary beds provide a stable optical pathlength, they are sites of active metabolism, so determining total Hgb concentration requires technique tolerances for large variations in oxygen saturation. Fortunately, techniques such as those disclosed above can overcome this limitation because oxy- and deoxyhemoglobin have numerous isosbestic points in the visible spectral range, allowing blood vessels to be imaged with visible light. Although many imaging studies have shown excellent performance in monitoring blood components, one potential clinical limitation of this class of device is the associated imaging instrumentation. High-resolution microvascular imaging requires focal plane arrays that may be expensive and have considerable associated electronics, making these devices cost-ineffective compared to standard invasive blood testing procedures. Still, with advancement in detector size, speed, and cost, rapid anemia screening tools using microvascular imaging can be implemented. This class of device may have maximum usefulness as a bedside stand-alone device rather than a mobile device. nir transmission spectroscopy NIR transmission spectroscopy is the embodiment of choice in the majority of studies of total Hgb detection, particularly methods of transcutaneous illumination of the fingertip. The primary emphasis for investigating transcutaneous Hgb monitoring techniques is the potential for integration into existing pulse oximeters, making these devices fully functional to measure pulse Hgb saturation (SpO2) as well as total Hgb concentration. Challenges of measuring total Hgb concentration, compared to relative measurements such as SpO2, lie in the contributions of other skin chromophores, variations in blood vessel location and density, variation in spectral signatures of oxy- and deoxyhemoglobin in the NIR regime, and changes in vessel diameter and subsequent optical pathlength during pulsation. Nevertheless, investigators have overcome these challenges and fabricated noninvasive Hgb screening devices utilizing transcutaneous illumination. Aldrich et al. have reported on the ability to use NIR transmission through the fingertip at a single pseudoisosbestic wavelength (905 nm) coupled with a sonomicrometer to monitor pulsatile changes in the optical pathlength through the finger as well as correct for interpatient variation in finger diameter (25). Sonomicrometers were positioned at opposite sides of the finger to monitor the pathlength fluctuations during systolic pulsation, and subsequently used to normalize transmission of NIR irradiation, as shown in Fig. 2 . Using a linear regression model, this group obtained a cross-validation correlation r = 0.84 (n = 24) for Hgb compared with a Coulter cell counter, with a sensitivity and specificity of 94% and 78%, respectively, for detecting anemia, and a mean error of prediction of 1.1 g/dL for Hgb. While providing accurate estimations of the pulsatile pathlength change, transducer-mediated monitoring makes integration into a pulse oximeter less straightforward, especially considering the need to secure the transducers on the skin using adhesives or other fluids. Figure 2. Open in new tabDownload slide Setup of transcutaneous illumination coupled with pathlength modulation detection using sonomicrometer transducers (A) and the detected variation in light intensity corresponding to the sonomicrometer detected variation in optical pathlength (B). Adapted and reproduced with permission from (25). A.U., absorbance units. Figure 2. Open in new tabDownload slide Setup of transcutaneous illumination coupled with pathlength modulation detection using sonomicrometer transducers (A) and the detected variation in light intensity corresponding to the sonomicrometer detected variation in optical pathlength (B). Adapted and reproduced with permission from (25). A.U., absorbance units. A wholly optical method for direct measurement of Hgb noninvasively was reported by Jeon et al., who used a 5-wavelength diode-emitting array (26). In diffuse photon propagation through the fingertip, a segment of irradiation photons will interact with the arterial vessels while another segment will encounter only soft tissue of the fingertip. Signals reaching the photodetector that passes through the arterial vessel are subject to pulsatile intensity modulation as a result of a varying vessel diameter and erythrocyte interactions with each heartbeat, while the signal bypassing the arterial vessel remains constant with pulsatile waveforms, enabling isosbestic wavelengths passing through the arterial vessel to be pathlength normalized and used to estimate Hgb. This group reported a cross-validation correlation of r = 0.804 (n = 97) for a calibration set and relative prediction error of 8.5% (n = 32) for a prediction set compared to reference Hgb obtained with the potassium cyanide method. Although this method requires more robust detection mechanisms for photons diffusing through the fingertip, it can be implemented quite easily in contemporary clinical pulse oximeters. Volumetric restriction of blood flow to the finger and monitoring of changes in transmission is an attractive method for differentiating the absorption contributions of Hgb/blood from those of surrounding tissue chromophores and therefore improving accuracies over purely optical methods. Rendell et al. have explored NIR transmission through the fingertip as a marker of total Hgb concentration (as well as blood flow) while stabilizing the finger and modulating blood flow using volume constrictive finger inserts (27). Applying a linear regression algorithm to multiple discrete spectral points taken both with and without the constrictive insert, this group obtained an optimum nonlinear regression correlation of r = 0.862 (n = 121) between radiation intensity values of multiple wavelengths and Hgb reference determined by the potassium cyanide method. Geva et al. examined a similar technique termed occlusion spectroscopy, which also uses fingertip NIR transmission and blood flow modulation in a commercial instrument (NBM-100, Orsense) (28). The transmission signal is varied by occluding blood flow with a restrictive finger cuff (similar to a blood pressure cuff), accelerating ischemia-driven RBC aggregation. The occluding finger cuff is later released, allowing RBC to disaggregate while the change in optical transmission at multiple NIR/visible wavelengths is monitored (28). Clinical evaluation of this commercial technology has shown average error values of 0.78 g/dL (n = 110) using predefined calibrations for Hgb compared to an undisclosed in vitro method (29). Powell et al. report on the accuracy of a similar device (Critscan, Hemametrics) that also uses a finger cuff to modulate the volume of blood flow into the finger and measures transcutaneous optical absorptions to infer Hct, with a correlation of r = 0.88 (n = 121), again using predefined calibrations for Hct compared to capillary-centrifuged in vitro tests (30). Although NIR transcutaneous methods of Hgb measurement have been thoroughly studied, they have yet to gain widespread clinical adoption. This situation may be partly a function of the technology adoption barrier for medical devices, and underlying technological barriers may also be prohibiting clinical use. As highlighted earlier, limitations in this method can include interference from the many other chromophores in skin and large variations in photon interaction lengths from varying tissue thicknesses. Of the reported methods discussed above, the flow modulation technique should be the most tolerant of these variations, because the only parameter being modulated during occlusion and release is Hct/Hgb, allowing for normalization and removal of interference artifacts. Consequently, this class of device has appeared in commercially manufactured instruments and has the potential for more widespread clinical adoption in the future. reflectance spectroscopy Several studies have investigated the use of reflectance spectroscopy as an alternative method to measure Hgb/Hct. An advantage of reflectance spectroscopy is that features close to the surface are probed, so deep transmission through highly scattering turbid tissue is not necessary. Because the required optical penetration depths are considerably lower, shorter wavelength light can be used, eliminating much of the variation between oxy and deoxyhemoglobin. In one such study, Wu et al. examined steady-state visible and NIR diffuse reflectance spectroscopy from the dorsal side of the arm and its correlation to Hgb and Hct (31). A linear regression analysis of the absorbance data resulted in cross-validation correlation r = 0.8 and an SE of cross validation of 0.9 g/dL for Hgb compared to a photometric reference instrument. Furthermore, this group used an in-house fabricated fiber-based temperature controlled reflectance spectrophotometer with six 10-nm bandpass filters and associated photodiodes for data collection. Temperature control over a small depth in the tissue is attractive because absorption/scattering coefficients and cutaneous blood flow have been shown to vary with local temperature variation (32), contributing to an unwanted variability in noninvasive Hgb measurement. Using a Montecarlo simulation and partial least squares analysis, Wu et al. obtained a cross-validation correlation of r = 0.8 and an SE of cross validation of 0.8 g/dL for Hgb (again compared to photometric reference method) in a patient set with 10 light-skinned patients tested multiple times (n = 26). Zhang et al. reported on the utility of a similar visible/NIR diffuse reflectance spectroscopy technique (33). A fiber optic probe was attached against the patient’s forearm (34), and NIR reflectance spectra were collected during cardiopulmonary bypass surgery. Because blood is diluted through heart-lung machines during cardiopulmonary bypass, dynamic intrapatient Hct concentrations can be evaluated over the course of the procedure. Using partial least squares regression, this group reported obtained an intrapatient cross-validation correlation of r = 0.844 (n = 10) between optical methods and a reference capillary centrifuge method, and an interpatient cross-validation correlation of r = 0.509 assessed by comparing cross-patient data using a single partial least squares regression model. This group has reported variations of this technology in additional human patient studies (35) as well as animal models (36). The studies listed above examine visible/NIR reflectance spectra from thick tissue regions and are consequently susceptible to variability in patients of different ethnicity as a result of melanin fluctuation, particularly in the short-wave visible regime. To minimize this error, a calibration should be built into the technique to normalize contributions from melanin, or the use of mucosal surfaces should be explored. McMurdy et al. reported on the use of visible reflectance spectroscopy to analyze the mucosal palpebral conjunctiva (inner lining of the eyelid) as a method of noninvasively monitoring Hgb (37). In a nonlinear regression model from diffuse reflectance spectra from the palpebral conjunctiva, a cross-validation correlation of r = 0.92 (n = 30) was obtained for Hgb compared to CBC testing across an ethnically diverse population. Two additional studies have shown that the use of either a standardized color chart (38) or photographs (39) of the conjunctiva as a comparative tool for the patient’s palpebral conjunctiva hue improved the sensitivity and specificity of observational methods. Although reflectance techniques may be more difficult to implement in existing medical devices, their comparable if not improved accuracy over transmission methods may enable their adoption as a new, stand-alone clinical screening tool. Like transcutaneous methods, devices using reflectance spectroscopy may be fabricated inexpensively and compactly, giving this class of device the same point-of-care capability across a wide scope. ultrasound and optoacoustic spectroscopy The dependence of ultrasonic wave generation and propagation through tissue on blood constituents leads to a completely different set of technologies to noninvasively monitor Hct/Hgb, including optoacoustics and pulse-echo ultrasound. Optoacoustic spectroscopy has been explored as a modality for noninvasive Hgb measurement in multiple in vitro (40) and in vivo (41)(42) studies. The rapid thermal expansion of the tissue through laser absorption creates an optoacoustic (pressure) wave, the characteristics of which depend on the characteristics of the absorbing analyte(s). In the NIR regime, Hgb has a higher absorption coefficient than surrounding tissue, enhancing optically induced thermal modulation. Fig. 3A shows the variation in optoacoustic signal with Hgb concentration in a tissue phantom. The superficial radial artery is an effective location for optoacoustic stimulation in the NIR because the vessel is close to the surface and the saturation is approximately 100%, eliminating errors caused from the variation in absorption spectra from oxy- and deoxyhemoglobin. Optoacoustic waves generated from the melanin layers in skin can be distinguished from those generated from the radial artery based on the phase delay in the pressure wave (41). Fig. 3B shows the optoacoustic signals coming from the melanin layers vs radial artery and the variation in this signal with saline infusion. The peak-to-peak intensity from the optoacoustic waves measured during in vivo dilution closely follows those found by measuring Hgb concentration directly in the in vitro dilution experiment, although no interor intrapatient predictive correlation was reported (42). Figure 3. Open in new tabDownload slide Modulation in optically generated acoustic wave as a function of Hgb concentration in a radial artery tissue phantom (A) and temporal variation in optoacoustic signal generated in the radial artery during saline infusion (B). Adapted from (42). Reproduced with permission. Figure 3. Open in new tabDownload slide Modulation in optically generated acoustic wave as a function of Hgb concentration in a radial artery tissue phantom (A) and temporal variation in optoacoustic signal generated in the radial artery during saline infusion (B). Adapted from (42). Reproduced with permission. Along similar lines, Secomski et al. have discussed the use of pulse echo ultrasound in determining Hct (43). Specifically, Hct is determined by monitoring the ultrasonic wave attenuation in blood, calculated from the Doppler power spectrum. Investigation of both a multigate Doppler system (44) and a single transducer pulse echo device verified the predictive ability of this method for Hct in an in vitro animal blood model and in an in vivo human patient set. Secomski reported an optimal correlation coefficient of r = 0.90 (n = 168) for an in vitro study using the pulse echo method and an optimal correlation coefficient of r = 0.96 (n = 14) for an in vitro study using a Doppler gating system (43). In the 6-patient in vivo study, again using the gated Doppler method, Hct determined from the brachial artery was within 5% of the value obtained with capillary centrifugation but Hct determined from the carotid artery was only within 20%. Although these studies of these methods are interesting, it is not clear how they will compare to optical methods. In their current configurations, optoacoustic and ultrasonic methods of Hgb/Hct determination are more difficult to implement in a clinical setting than spectroscopic or imaging methods, inhibiting their application in numerous settings discussed above. Further evaluation and clinical trials of these techniques will indicate the potential of commercially viable technologies. Conclusions Among the new methods described, the best have shown accuracy of 95% relative to in vitro measured Hgb/Hct. Although it is difficult at this point to compare the overall performance of each class of technique, general comments on the results of these studies seem to show that (a) more complex techniques using conductance and optoacoustics can have excellent performance if suitable for the particular environment, (b) transcutaneous optical measurements yield better results with blood flow modification to reject interference from background tissue properties, (c) for the same tissue interference reasons, optical reflectance measurements do not perform as well, and (d) exploration of mucosal surfaces through imaging and spectroscopy is an additional way to reject tissue interference. Table 1 shows correlations in multipatient trials. Table 1. Correlation coefficients for the multipatient studies evaluated in this review. Correlation coefficients were determined with cross-validation, calibration and measurement of a prediction set, or correlation to a defined parameter. Author . Technique . Sample size . r value . Calibration method . Yamakoshi 1980 Electrical admittance plethysmography of finger 16 0.98 (Hct) In vitro calibration Kanashima 2005 NIR transcutaneous imaging (finger) 140 0.59 (Hgb) Prediction set Nadeau 2001 Orthogonal polarization imaging 71 0.93 (Hgb) Prediction set Rice 2001 Retinal spectral imaging 24 0.89 (Hgb) Cross validation Iftimia 2006 Retinal tracking and low-coherence interferometry 7 Not reported Not applicable Aldrich 2002 Photoplethymography (finger) 24 0.84 (Hgb) Cross validation Jeon 2002 NIR transcutaneous spectroscopy (finger) 129 0.80 (Hgb) Cross validation Rendell 2003 NIR transcutaneous spectroscopy (finger) 121 0.86 (Hgb) Nonlinear regression correlation Geva 2004 NIR occlusion spectroscopy (finger) 110 Not reported Not applicable Powell 2005 NIR transcutaneous spectroscopy (finger) 121 0.88 (Hct) Prediction set Wu 2000 Reflectance spectroscopy (forearm) 28 0.80 (Hgb) Cross validation Zhang 2000 Reflectance spectroscopy (forearm) 10 0.51 (Hct) Cross validation McMurdy 2006 Reflectance spectroscopy (conjunctiva) 30 0.92 (Hgb) Cross validation Author . Technique . Sample size . r value . Calibration method . Yamakoshi 1980 Electrical admittance plethysmography of finger 16 0.98 (Hct) In vitro calibration Kanashima 2005 NIR transcutaneous imaging (finger) 140 0.59 (Hgb) Prediction set Nadeau 2001 Orthogonal polarization imaging 71 0.93 (Hgb) Prediction set Rice 2001 Retinal spectral imaging 24 0.89 (Hgb) Cross validation Iftimia 2006 Retinal tracking and low-coherence interferometry 7 Not reported Not applicable Aldrich 2002 Photoplethymography (finger) 24 0.84 (Hgb) Cross validation Jeon 2002 NIR transcutaneous spectroscopy (finger) 129 0.80 (Hgb) Cross validation Rendell 2003 NIR transcutaneous spectroscopy (finger) 121 0.86 (Hgb) Nonlinear regression correlation Geva 2004 NIR occlusion spectroscopy (finger) 110 Not reported Not applicable Powell 2005 NIR transcutaneous spectroscopy (finger) 121 0.88 (Hct) Prediction set Wu 2000 Reflectance spectroscopy (forearm) 28 0.80 (Hgb) Cross validation Zhang 2000 Reflectance spectroscopy (forearm) 10 0.51 (Hct) Cross validation McMurdy 2006 Reflectance spectroscopy (conjunctiva) 30 0.92 (Hgb) Cross validation Open in new tab Table 1. Correlation coefficients for the multipatient studies evaluated in this review. Correlation coefficients were determined with cross-validation, calibration and measurement of a prediction set, or correlation to a defined parameter. Author . Technique . Sample size . r value . Calibration method . Yamakoshi 1980 Electrical admittance plethysmography of finger 16 0.98 (Hct) In vitro calibration Kanashima 2005 NIR transcutaneous imaging (finger) 140 0.59 (Hgb) Prediction set Nadeau 2001 Orthogonal polarization imaging 71 0.93 (Hgb) Prediction set Rice 2001 Retinal spectral imaging 24 0.89 (Hgb) Cross validation Iftimia 2006 Retinal tracking and low-coherence interferometry 7 Not reported Not applicable Aldrich 2002 Photoplethymography (finger) 24 0.84 (Hgb) Cross validation Jeon 2002 NIR transcutaneous spectroscopy (finger) 129 0.80 (Hgb) Cross validation Rendell 2003 NIR transcutaneous spectroscopy (finger) 121 0.86 (Hgb) Nonlinear regression correlation Geva 2004 NIR occlusion spectroscopy (finger) 110 Not reported Not applicable Powell 2005 NIR transcutaneous spectroscopy (finger) 121 0.88 (Hct) Prediction set Wu 2000 Reflectance spectroscopy (forearm) 28 0.80 (Hgb) Cross validation Zhang 2000 Reflectance spectroscopy (forearm) 10 0.51 (Hct) Cross validation McMurdy 2006 Reflectance spectroscopy (conjunctiva) 30 0.92 (Hgb) Cross validation Author . Technique . Sample size . r value . Calibration method . Yamakoshi 1980 Electrical admittance plethysmography of finger 16 0.98 (Hct) In vitro calibration Kanashima 2005 NIR transcutaneous imaging (finger) 140 0.59 (Hgb) Prediction set Nadeau 2001 Orthogonal polarization imaging 71 0.93 (Hgb) Prediction set Rice 2001 Retinal spectral imaging 24 0.89 (Hgb) Cross validation Iftimia 2006 Retinal tracking and low-coherence interferometry 7 Not reported Not applicable Aldrich 2002 Photoplethymography (finger) 24 0.84 (Hgb) Cross validation Jeon 2002 NIR transcutaneous spectroscopy (finger) 129 0.80 (Hgb) Cross validation Rendell 2003 NIR transcutaneous spectroscopy (finger) 121 0.86 (Hgb) Nonlinear regression correlation Geva 2004 NIR occlusion spectroscopy (finger) 110 Not reported Not applicable Powell 2005 NIR transcutaneous spectroscopy (finger) 121 0.88 (Hct) Prediction set Wu 2000 Reflectance spectroscopy (forearm) 28 0.80 (Hgb) Cross validation Zhang 2000 Reflectance spectroscopy (forearm) 10 0.51 (Hct) Cross validation McMurdy 2006 Reflectance spectroscopy (conjunctiva) 30 0.92 (Hgb) Cross validation Open in new tab Although reports reviewing techniques were first published more than 2 decades ago, a truly noninvasive device has yet to be adopted as a standard of care. The likely reasons for the lack of physician adoption include the complicated nature of data collection, necessary instrumentation, and inaccuracy/repeatability issues. These difficulties have not prevented numerous commercial devices from appearing in clinical studies, such as those used and evaluated in the preceding text. Efficient penetration of this technology hinges on the simplification and miniaturization of these devices so they can be implemented easily into patient care while maintaining clinically acceptable performance. Grant/funding Support: The authors wish to acknowledge their own research funding support from the NASA GSRP Fellowship (NNG05GL57H), NCIIA Advanced E-Team Program, and the Goldman Philanthropic Charles E. Culpeper Biomedical Initiative Pilot Program. Financial Disclosures: J.W. McMurdy, G.P. Crawford, S. Suner, and G.D. Jay are all cofounders of Corum Medical, Inc. Acknowledgments: The authors wish to acknowledge assistance in manuscript preparation from members of the Display and Photonics laboratory at Brown University. 1 " Nonstandard abbreviations: RBC, red blood cells; Hgb, hemoglobin; Hct, hematocrit; CBC, complete blood count; NIR, near-infrared; OPS, orthogonal polarization spectral; SpO2, pulse saturation. References 1 Goodnough LT, Nissenson AR. Anemia and its clinical consequences in patients with chronic diseases. Am J Med 2004 ; 116 : 1 -2. Crossref Search ADS PubMed 2 Nissenson AR, Goodnough LT, Dubois RW. Anemia: not just an innocent bystander?. Arch Intern Med 2003 ; 163 : 1400 -1404. Crossref Search ADS PubMed 3 Verhoef H, West CE, Kraaijenhagen R, Nzyuko SM, King R, Mbandi MM, et al. Malarial anemia leads to adequately increased erythropoiesis in asymptomatic Kenyan children. Blood 2002 ; 100 : 3489 -3494. Crossref Search ADS PubMed 4 Ulijaszek SJ. Between-population variation in preadolescent growth. Eur J Clin Nutr 1994 ; 48 : S5 -S14. PubMed 5 . World Health Organization. Nutrional anemias: report of the WHO Scientific Group. WHO Technical Report Series 1968 : 415 . 6 DeMaeyer E, Adiels-Yagman M. The prevalence of anemia in the world. World Health Statistics 1985 ; 38 : 302 -316. 7 Toto RD. Anemia of chronic disease: past, present, and future. Kidney Int 2003 ; 64 : S20 -SS3. Crossref Search ADS 8 Pereira AA, Sarnak MJ. Anemia as a risk factor for cardiovascular disease. Kidney Int 2003 ; 64 : S32 -S39. Crossref Search ADS 9 National Anemia Action Council. The prevalence and impact of anemia: a systematic review of the published literature.http://www.anemia.org/about_anemia/systematic_review.jsp (Accessed 2005). 10 Hung OL, Kwon NS, Cole AE, Dacpano GR, Wu T, Chiang WK, Goldfrank LR. Evaluation of the physician’s ability to recognize the presence or absence of anemia, fever, and jaundice. Acad Emer Med 2000 ; 7 : 146 -156. Crossref Search ADS 11 Kost GJ, Tran NK, Tuntideelert M, Kulrattanamaneeporn S, Peungposop N. Katrina, the tsunami, and point-of-care testing: optimizing rapid response diagnosis in disasters. Am J Clin Pathol 2006 ; 126 : 513 -520. Crossref Search ADS PubMed 12 Yamakoshi KI, Shimazu H, Togawa T, Fukuoka M, Ito H. Non-invasive measurement of hematocrit by electrical admittance plethysmography technique. IEEE Trans Biomed Eng 1980 ; 27 : 156 -161. PubMed 13 Yamakoshi K, Tanaka S, Shimazu H. Electrical admittance cuff for noninvasive and simultaneous measurement of hematocrit, arterial-pressure and elasticity using volume-oscillometric method. Med Biol Eng Comput 1994 ; 32 : S99 -S107. Crossref Search ADS PubMed 14 Kanashima H, Yamane T, Takubo T, Kamitani T, Hino M. Evaluation of noninvasive hemoglobin monitoring for hematological disorders. J Clin Lab Anal 2005 ; 19 : 1 -5. Crossref Search ADS PubMed 15 Saigo K, Imoto S, Hashimoto M, Mito H, Moriya J, Chinzei T, et al. Noninvasive monitoring of hemoglobin: the effects of WBC counts on measurement. Am J Clin Pathol 2004 ; 121 : 51 -55. Crossref Search ADS PubMed 16 Nadeau RG, Groner W. The role of a new noninvasive imaging technology in the diagnosis of anemia. J Nutr 2001 ; 131 : 1610S -1614S. Crossref Search ADS PubMed 17 Genzel-Boroviczeny O, Harris AG, Fletcher GD, Christ F. Noninvasive determination of hemoglobin in neonates with orthogonal polarization spectral (OPS) imaging. Pediatr Res 2000 ; 47 : 332A-A . 18 Winkelman JW. Noninvasive blood bell measurements by imaging of the microcirculation. Am J Clin Pathol 2000 ; 113 : 479 -483. Crossref Search ADS PubMed 19 Rice MJ, Sweat J, R. H., Rioux JM, Williams WT, Routt W. Non-invasive measurement of blood components using retinal imaging. United States Patent 2002 : 6477394 . 20 Winkelman JW. Apparatus and method for in vivo analysis of red and white blood cells. United States Patent 1991 : 4998533 . 21 Iftimia NV, Hammer DX, Bigelow CE, Rosen DI, Ustun T, Ferrante AA, et al. Toward noninvasive measurement of blood hematocrit using spectral domain low coherence interferometry and retinal tracking. Opt Exp 2006 ; 14 : 3377 -3388. Crossref Search ADS 22 Leitgeb R, Hitzenberger CK, Fercher AF. Performance of Fourier domain vs. time domain optical coherence tomography. Opt Exp 2003 ; 11 : 889 -894. Crossref Search ADS 23 Faubert J, Gagnon M, Diaconu V. The OSOME system: a new real-time non-invasive technique for measuring oxygenation and hemoglobin in the human eye. Invest Ophth Vis Sci 1999 ; 40 : S977-S . 24 Diaconu V, Faubert J. On-line and real-time spectroreflectometry measurement of oxygenation in a patient’s eye. United States Patent 2000 : 6149589 . 25 Aldrich TK, Moosikasuwan M, Shah SD, Deshpande KS. Length-normalized pulse photoplethysmography: a noninvasive method to measure blood hemoglobin. Ann Biomed Eng 2002 ; 30 : 1291 -1298. Crossref Search ADS PubMed 26 Jeon KJ, Kim SJ, Park KK, Kim JW, Yoon G. Noninvasive total hemoglobin measurement. J Biomed Opt 2002 ; 7 : 45 -50. Crossref Search ADS PubMed 27 Rendell M, Anderson E, Schlueter W, Mailliard J, Honigs D, Rosenthal R. Determination of hemoglobin levels in the finger using near infrared spectroscopy. Clin Lab Haematol 2003 ; 25 : 93 -97. Crossref Search ADS PubMed 28 Shvartsman LD, Fine I. Optical transmission of blood: effect of erythrocyte aggregation. IEEE Trans Biomed Eng 2003 ; 50 : 1026 -1033. Crossref Search ADS PubMed 29 Fine I, Geva D, Amir O, Monashkin E. Clinical Evaluation of non-invasive hemoglobin testing device. American Society of Anesthesiologists Annu Meeting Vol. A-597. New Orleans, LA 2004 . 30 Powell JS, Gresens CJ, Miller D, Dummer W, Holland P. Non-invasive measurement of hematocrit to screen potential blood donors. Transfusion 2005 ; 45 : 157A-A . 31 Wu XM, Yeh S, Jeng TW, Khalil OS. Noninvasive determination of hemoglobin and hematocrit using a temperature-controlled localized reflectance tissue photometer. Anal Biochem 2000 ; 287 : 284 -293. Crossref Search ADS PubMed 32 Khalil OS, Yeh S, Lowery MG, Wu X, Hanna CF, Kantor S, et al. Temperature modulation of the visible and near infrared absorption and scattering coefficients of human skin. J Biomed Opt 2003 ; 8 : 191 -205. Crossref Search ADS PubMed 33 Zhang SB, Soller BR, Kaur S, Perras K, Vander Salm TJ. Investigation of noninvasive in vivo blood hematocrit measurement using NIR reflectance spectroscopy and partial least-squares regression. Appl Spectrosc 2000 ; 54 : 294 -299. Crossref Search ADS 34 Soller BR, Cabrera M, Smith SM, Sutton JP. Smart medical systems with application to nutrition and fitness in space. Nutrition 2002 ; 18 : 930 -936. Crossref Search ADS PubMed 35 Soller BR, Zhang S, Kaur S, Heard SO, Gaca C, Rohrer MJ, et al. Application of near infrared spectroscopy for the noninvasive measurement of hematocrit on surgical patients. Crit Care Med 1999 ; 27 : A144-A . 36 Soller BR, Cingo N, Puyana JC, Khan T, Hsi C, Kim H, et al. Simultaneous measurement of hepatic tissue pH, venous oxygen saturation and hemoglobin by near infrared spectroscopy. Shock 2001 ; 15 : 106 -111. Crossref Search ADS PubMed 37 McMurdy JW, Jay GD, Suner S, Trespalacios FM, Crawford GP. Diffuse reflectance spectra of the palpebral conjunctiva and its utility as a noninvasive indicator of total hemoglobin. J Biomed Opt 2006 ; 11 : 014019 . Crossref Search ADS PubMed 38 Shapiro NI, Stumpf M, Chiota N, Resciniti M, Nicolet J, Ufberg J. A standardized conjunctiva pallor tool predicts anemia. Acad Emerg Med 2004 ; 11 : 478 . 39 Ernsting K, Suner S, Jay GD. Use of digital imaging of conjunctiva to predict hemoglobin concentration. Acad Emerg Med 2001 ; 8 : 528 -529. 40 Deyo DJ, Esenaliev RO, Hartrumpf O, Motamedi M, Prough DS. Continuous noninvasive optoacoustic monitoring of hemoglobin concentration. Anesthesiol Analgesia 2001 ; 92 : S139 . 41 Esenaliev RO, Petrov YY, Hartrumpf O, Deyo DJ, Prough DS. Continuous, noninvasive monitoring of total hemoglobin concentration by an optoacoustic technique. Appl Opt 2004 ; 43 : 3401 -3407. Crossref Search ADS PubMed 42 Petrova IY, Esenaliev RO, Petrov YY, Brecht HPE, Svensen CH, Olsson J, et al. Optoacoustic monitoring of blood hemoglobin concentration: a pilot clinical study. Opt Lett 2005 ; 30 : 1677 -1679. Crossref Search ADS PubMed 43 Secomski W, Nowicki A, Guidi F, Tortoli P, Lewin PA. Noninvasive in vivo measurements of hematocrit. J Ultrasound Med 2003 ; 22 : 375 -384. Crossref Search ADS PubMed 44 Liang RI, Prapas N, Detti L, Cosmi E, Copel JA, Mari G. Assessment of blood flow velocity waveforms of the pulmonary circulation by multigate spectral Doppler scanning and traditional pulsed Doppler ultrasonography. J Ultrasound Med 2002 ; 21 : 31 -37. Crossref Search ADS PubMed © 2008 The American Association for Clinical Chemistry This article is published and distributed under the terms of the Oxford University Press, Standard Journals Publication Model (https://academic.oup.com/journals/pages/open_access/funder_policies/chorus/standard_publication_model) TI - Noninvasive Optical, Electrical, and Acoustic Methods of Total Hemoglobin Determination JF - Clinical Chemistry DO - 10.1373/clinchem.2007.093948 DA - 2008-02-01 UR - https://www.deepdyve.com/lp/oxford-university-press/noninvasive-optical-electrical-and-acoustic-methods-of-total-0pbwOAb9sb SP - 264 VL - 54 IS - 2 DP - DeepDyve ER -