TY - JOUR AU - Lyons, Norman AB - A method is described for the determination of GFR and ERPF using a combined injection of 51Cr-EDTA and 125I-iodohippurate. Plasma samples obtained at 60 min and 150 min after administration were used to determine a flow rate F assuming a monoexponential clearance of the tracers. Empirical relationships were found between F and the true GFR and ERPF determined from multiple sampling and multi-exponential analysis of the clearance curves. The method was shown to be superior to previously published methods involving one or two samples. GRF was calculated with a standard deviation (SD) of only 3.3 ml/min when compared to the multiple sample technique (the reference method), whereas the best estimate from a single sample had a SD of 6.3 ml/min. An improvement in accuracy of ERPF estimation was achieved for values of flow of less than 200 ml/min when the SD was only 9.3 ml/min. TI - Simplified estimation of glomerular filtration rate and effective renal plasma flow JF - European Journal of Nuclear Medicine and Molecular Imaging DO - 10.1007/BF00251347 DA - 2004-08-05 UR - https://www.deepdyve.com/lp/springer-journals/simplified-estimation-of-glomerular-filtration-rate-and-effective-0WaIxK4n9M SP - 249 EP - 253 VL - 6 IS - 6 DP - DeepDyve ER -